S P E C I F I C A T I O N S
8.7 Calibration
Test with factory calibration software to verify that USB-SA44B is operating within tolerance.
Recommended calibration interval is 1 year at 20°C to 25°C.
8.8 Adjustments
Factory adjustment software shall be used to generate new calibration constants when USB-SA44B
will not pass calibration. Temperature correction data is generated only at time of manufacture.
8.9 Optional Accessories
1) 20dB Fixed Attenuator, Mini Circuits Part # VAT-20+
2) DC Block, Mini Circuits Part # BLK-89+
3) SMA(m) to BNC(f) Adapter, Mini Circuits part# SM-BF50+
4) 3GHz RF Probe, p/n P-20A (made by Auburn Technology,
www.auburntec.com)
35