Specifications; Frequency; Amplitude (Rbw ≤100Khz) - Test Equipment Plus Signal Hound USB-SA44B User Manual

Usb-powered, spectrum, 1hz to 4.4ghz
Table of Contents

Advertisement

S P E C I F I C A T I O N S

8 Specifications

Unless otherwise stated, specifications are valid for an ambient temperature range of 0
to 70°C (or -40 to +85°C for option 1), image rejection on, amplitude of signal applied
less than the reference level.

8.1 Frequency

Frequency Range: 1 Hz to 4.4 GHz
Span Modes: (Center Frequency + Span) or (Start + Stop Frequencies)
Maximum Span: 4.4 GHz
Minimum Span: 10 Hz or Zero Span
Internal Frequency Reference Accuracy: ±1 ppm
For greater frequency accuracy connect to external 10 MHz timebase (0 to +20 dBm)
Frequency Readout Accuracy: reference error ±1 sample
Marker Accuracy: reference error ±1 sample
Resolution Bandwidth: 0.1Hz to 250KHz and 5MHz
Spectral Purity
Residual FM, 3KHz Audio LPF, 15 KHz IF BW: [0.1 Hz + 4 Hz / GHz] typical RMS
FM (e.g. 2 GHz RF would have 8.1 Hz RMS FM). Increasing IF BW increases residual
FM.
Note 1: 1 sample represents approximately 40% of the selected RBW
Note 2: 5 MHz RBW accuracy not specified, software support coming soon
8.2
Amplitude (RBW ≤100KHz)
Range: 1dB Gain Compression to Displayed Average Noise Level (DANL)
1dB Gain Compression: (attenuator set to 15dB, preamp off)
+16dBm Typical, 1Hz to 150MHz
+19dBm Typical, 150MHz to 4.4GHz
Displayed Average Noise Level: 0dB input attenuation, 1Hz RBW
Frequency
10Hz
100Hz to 10KHz
10KHz to 500KHz
500KHz to 10MHz
10MHz to 100MHz
100MHz to 1GHz
1GHz to 2.6GHz
2.6GHz to 3.3GHz
3.3GHz to 4.4GHz
Absolute Accuracy (Reference level ≤0 dBm): ±1.5dB, no warm-up required
Absolute Accuracy (0 dBm < Reference level ≤ 10 dBm): ±2.0dB
(2)
RF Preamp Off
RF Preamp On
-124 dBm
NA
-130 dBm
NA
-142 dBm
NA
-142 dBm
-153 dBm
-148 dBm
-161 dBm
-144 dBm
-158 dBm
-139 dBm
-151 dBm
-135 dBm
-151 dBm
-128 dBm
-134 dBm