Absolute Maximum Ratings - AEA VIA Bravo MRI Operating Manual

Mri network analyzer mri network analyzer - 100khz to 70 mhz
Table of Contents

Advertisement

9.1.4.2
Serial Port
The serial port allows a PC or other terminal to control and read the VIA
Bravo MRI unit. Use of VIA Director Software is highly recommended.
Connect VIA Bravo MRI to PC with DB9F to 1/8 inch stereo plug serial port
cable AEA PN 0070-1201.
9.1.4.2.1
Bit Rates
Bit Rate selectable: 4800, 9600, 19200, 28800, 57600
9.1.4.2.2
Data Format
Data type 8 bit, no parity, and 1 stop bit, with XON/XOFF handshake
9.1.4.2.3
Connector Type
Serial input is a 1/8 inch stereo jack. Tip = RD (from VIA Bravo MRI to
PC), Ring = TD (from PC to VIA Bravo MRI), Common=ground
9.1.4.3
Power Requirements
Using the AEA wall power pack (PN 5001-0201) is recommended to
supply the DC input to the VIA Bravo MRI unit.
9.1.4.3.1 DC input (Power jack)
Min 12 VDC (300 ma)
Typical 14VDC (250ma)
Max 20VDC (250 ma)
Positive polarity is applied to the center pin of the power jack for proper
operation.
2.1 mm center pin DC power connector.
9.1.4.3.2 Battery Power
If battery power must be used, only use an external battery pack that will
sit outside the magnetic field. A sealed lead acid battery will avoid the
dangers of being pulled into the magnet, but you must still keep the battery
away from the field to prevent excessive discharge and heat. Operating a
battery too close to the magnet will present a safety hazard (internal heating).
The lead acid battery(s) should connect to the VIA Bravo MRI using the
power jack next to the RF connector. To recharge the battery(s), disconnect
the batteries from the VIA Bravo MRI.

9.1.5 Absolute Maximum Ratings

9.1.5.1.1 RF Input (RF jack)
+10dBm Max
+/- 16VDC Max
24