Due to the possibility of corrosion when used in certain environmental conditions, it is possible that the
product's safety could be compromised over time. It is important that the product be periodically inspected for
sound electrical connections and equipment grounding. Do not use if the equipment grounding or electrical
insulation has been compromised.

Safety Precautions - Service and operation

2.2
Ensure the enclosure of the VGC301 is connected directly to a good quality earth ground.
Ensure that the vacuum port on which the CVG101 vacuum gauge tube is mounted is electrically grounded.
Use an appropriate power source of 12 to 30 Vdc, 2 W or use InstruTech series PS301 optional power supplies.
Turn off power to the unit before attempting to service the controller.
Turn off power to the unit if a cable or plug is damaged or the product is not operating normally according to
this instruction manual. Contact qualified InstruTech service personnel for any service or troubleshooting
condition that may not be covered by this instruction manual.
It is important that the product be periodically inspected for sound electrical connections and equipment
grounding. Do not use if the equipment grounding or electrical insulation has been compromised.
Do not use if the unit has been dropped or the enclosure has been damaged. Contact InstruTech for return
authorization and instructions for returning the product to InstruTech for evaluation.

Electrical Conditions

2.3
WARNING! When high voltage is present in any vacuum system, a life threatening electrical shock hazard
may exist unless all exposed electrical conductors are maintained at earth ground potential. This applies to all
products that come in contact with the gas contained in vacuum chambers. An electrical discharge within a
gaseous environment may couple dangerous high voltage directly to any ungrounded conductor of electricity. A
person could be seriously injured or killed by coming in contact with an exposed, ungrounded electrical
conductor at high voltage potential. This condition applies to all products that may come in contact with the gas
inside the vacuum chamber (vacuum/pressure containment vessel).

2.3.1 Proper Equipment Grounding

WARNING! Hazardous voltages that could seriously injure or cause death are present in many vacuum
processes. Verify that the vacuum port on which the CVG101 vacuum gauge tube is mounted is electrically
grounded. Consult a qualified Electrician if you are in doubt about your equipment grounding. Proper
grounding of your equipment is essential for safety as well as intended operation of the equipment. The
CVG101 vacuum gauge tube and enclosure of the VGC301 controller must be connected directly to a good
quality earth ground. Use a ground lug on the CVG101 gauge vacuum connection / flange if necessary.
InstruTech, Inc.
VGC301 Controller
Page 7