Ohms - Datron 1061 Calibration And Servicing Handbook

Hide thumbs Also See for 1061:
Table of Contents

Advertisement

5
r'
F
OHMS
CALIBRATION TABLE
[1]
'Wilh
Standard Resistor
sourc€s
it
may
be
useful
to
use
the ,KEYBOARD'
method
of calibration
-
see
section
î.7
f2l
'For1061 A,
lnputfilterincreasesresolutionbyldigit,sol06lAfiguresaregiveninbrackets().
Remarks
Wait
for
the
reading
to
stabilize before
operating
'Zero'
Wait
for
the
reading
to
stabilize
before
operating 'Gain'
DVM
Reading
After
Calibration
f.0.0000s¿
t1
disit
10.0000s¿
i.1 digit
*.0.000ç¿
100.000ç¿
+..00000kç¿
1.00000ko
t0.0000ks¿
10.0000ko
f.0.000ko
100.000kQ
+.0.00kç¿
(+0.000kO)
1000.00ko
t1
digit
(1000.000ks¿)
(*10
¿ig¡ts)
È0.0000Mo
(*0.00000MQ)
10.0000Mo
.t5
digits
(10.00000Ms¿)
(iso
disits)
,CALIBRATE'
Key
'Zero'
'Gain'
'Zero'
'Gain'
'Zero'
'Gain'
'Zero'
'Gain'
Zero'
'Gain'
Zero'
'Gain'
'Zero'
'Gaín'
DVM
Setting
kO,4
wire,
10O
kO,4
wire, 10O
kO,4
wire,
.1
kS},4
wire,
.1
kl),4
wire,
1
kO,4
wíre,
1
kO,4 wire,
10
kO,4 wire,
10
kO,4 wire,
100
kO,4 wire,
100
kO,4 wire,
1000
lnput
Filterf
2l
kQ,4 wire,
1000
lnput F¡lter[2]
kO,4 wire,
1OMO
lnput
Filter[21
kO,4wire,lOMO
lnput
Filtert2l
Calibration
Source
4
wire
zero
10s¿
[1]
Standard
R
esistor
4 wire
zero
looo
[1]
Standard
Resistor
4
wire
zero
lko
u1
Standard
Resistor
4 wire
zero
loko
[1]
Standard
Resistor
4 wire
zero
100ko
[1]
Standard
Resistor
4 wire
zero
1000kç¿
[1]
Standard
Resistor
4
wire
zero
1oMO [1]
Standard
Resistor
Calibration
Operation
10Q
Range
Zero
10Q Full
Range
.1kQ
Range
Zero
lkO
Full
Range
1kO
Range
Zero
lkQ
Full
Range
1OkQ
Range
Zero
1OkO
Full
Range
100kQ
Range
Zero
100ko
Full
Range
1000ko
Range
Zero
1000ko
Full
Range
1OMO Range
Zero
1OMO
Full
Range
Step
1
2
3
4
5
6
7
I
9
10
11
12
13
14
; i
.r,
,:.,..';'a;.1;,1,.
.,,;:-.,
¡;..;;:¡.:.
. .,
.;
j:.j::jr;ji:iìr:::,:,..r,:......
: I :..
': ..
.: .. ri j tl'-lr:.
i-.
: ..-
.
.l