Connecting The Signal Cables - Liquid Controls LCI40 Series Installation And Parts

Signal converter for mass flowmeters
Table of Contents

Advertisement

4. eleCtriCal ConneCtionS

4.4 Connecting the Signal Cables

DANGER
Cables may only be connected
when the power is switched off.
DANGER
The device must be grounded
in accordance with regulations in order to
protect personnel against electric shocks.
DANGER
For devices used in hazardous
areas, additional safety notes apply; please
refer to the Ex documentation.
WARNING
Observe without fail the local
occupational health and safety regulations.
Any work done on the electrical components
of the measuring device may only be carried
out by properly trained specialists.
information
Each time a housing cover is
opened, the thread should be
cleaned and greased. Use only
resin-free and acid-free grease.
Ensure that the housing gasket
is properly fitted, clean and
undamaged.
16
4.4.1 Connection of Signal Cable - Field Housing and
Connection Box for Sensor
Connection of Signal Cable - Field Housing and Connection
Box for Sensor
1. Unscrew the terminal compartment cover.
2. Pass the prepared signal cable through the cable entry.
3. Secure the signal cable using the clip.
4. Connect the electrical conductors as shown. The shielding
5. Re-fit the cover and tighten it by hand.
Cable
Cable Pair
1
1
2
2
3
3
4
4
5
5
WWW.LCMETER.COM
must also be connected to the spring terminal.
Cable
Color
yellow
black
green
black
blue
black
red
black
white
black
Connection Terminal
X1 SA+
X1 SA-
X1 SB+
X1 SB-
X2 T1
X2 T2
X2 T3
X2 T4
X3 DR+
X3 DR-
LCI40
Series IOM