SECTION 8—TROUBLESHOOTING GUIDE
LCD DISPLAY:
DESCRIPTION:
Abnormal operating condition has been
detected
ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart.
Call your provider and report the problem if alarm continues.
LCD DISPLAY:
DESCRIPTION:
Battery pack is too hot, or too cold, to
allow battery to continue to operate.
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or
revert to normal operating screens
Invacare SOLO
2
INDICATORS:
Continuous audible beep every half
second.
RED indicator is On.
GREEN indicator is Off.
SOLUTIONS:
1. Turn unit Off, then On again to retry.
2. Change to another source of oxygen if
alarm continues.
INDICATORS:
Ten audible beeps every 10 sec
RED indicator Flashing Fast.
SOLUTIONS:
1. Remove battery and only use AC, or
DC, adapter for power. Or, use an
alternate battery pack.
2. Move battery to warmer, or cooler,
surroundings. Allow battery to cool down
to less than 95° F (35° C), or warm up to
50° F (10° C).
3. Turn unit off. Re-insert battery and
retry.
4. Change to another source of oxygen if
no other power source is available.
42