Viewing Individual Results
To see individual results for each wire pair, use
move among the screens.
In this mode, the tester continuously tests only the wire pair
you are viewing.
Figure 14 shows examples of these screens.
20
Short on pair 1,2 at 29.8 m.
or
; to
On the individual results screens, shorts are shown
only when they are between wires in a pair.
When there is a short, the far-end adapter and the
mapping of the unshorted wires are not shown.
Pair 3,6 is 67.7 m long and is terminated with the
wiremap adapter.
Open on pair 4,5 at 48.1 m. The open could be on one
or both wires.