2CT CLAMP LEAKER MODEL 2002 INSTRUCTION MANUAL Thank you very much for selecting our digital 2CT clamp leaker model 2002. This model is complex instrument and employ a very reliable mechanical/electronic design. Before you use your new instrument, read this instruction...
SAFETY SUMMARY observe by all means ● To use this instrument safely, read this “SAFETY SUMMARY” carefully and apply the instrument correctly. ● The CAUTIONs and WARNINGs which appear on the following pages are stated to prevent the operator & other people from the dangers and their properties from the damages beforehand.
SPECIFICATIONS Measuring Function: AC Line Current, Leakage Current, (2CT Method for Line & Leakage Current) AC Current Detection : True RMS Reading AD Conversion: Equivalent to Double Integration Method Display: 4 digit LCD max. reading of 9999 with annunciator Measuring Range: Leakage Current 0~1000mA/10A(50/60Hz) Line Current 0~20A/200A(50/60Hz) 2 CT Method 0~1000m0A/10A(50/60Hz)
OPERATION 〔Dimension and Panel Function〕 Sensor for detecting current and clamp method ①Clamp CT CT will open by pushing this lever to inside. ②Open/Close Lever To change power on/off and the ranges (2CT). ③Range Switch ④CT Selection Switch φ 5 for CTP-05-2 and φ 18 for ZCT-18-2 by 2CT method By pressing one time, DH (Data Hold) on LCD and it ⑤D-HOLD...
REPLACEMENT OF BATTERIES WARNING △ ! P OSSIBLE ELECTRICAL SHOCK OR ACCIDENT ● Do not replace the batteries under the conditions of measuring current or voltage ● Do not operate the instrument with battery cover off. CAUTION △ ! When not using the instrument for a long period, remove the batteries and keep separately.
MEASURING METHOD For the safety operation keep and pay attention to the cautions and warnings stated in this manual. WARNING △ ! P OSSIBLE ELECTRICAL SHOCK ● This instrument is for the use of low voltage circuit. Do not make measurements of power lines carrying more than AC 600V. Before use, check and confirm the voltage of circuit to be measured.
Page 7
(3) 2 CT Method Leakage Current Measurement measure leakage current where CT cannot applied conductors en block in the distribution circuit board like as a drawing on the right, by using spare CT (CTP-05-2). In this case, make instrument and spare CT to the same direction.
REPAIR SERVICE When requesting for repair service, please bring the instrument directly to the dealer where you bought. When mailing the instrument, always pack it in its original or equivalent packing materials to avoid any damage during the transportation and also put together with documents showing your name, address, phone number and defect point.