12.0 Routine (Operator) Cleaning
The BiliChek System is made of an easy-to-clean material
with a smooth, textured surface. One of the following clean-
ing agents is recommended for cleaning the exterior of the
HHU, Battery Charger Base, and Battery Packs:
KleenaSeptic
®
Cavicide
®
70 to 90% Isopropyl Alcohol solution
1% Bleach solution
Spray the cleaning agent on a soft damp cloth and wipe clean
the BiliChek system. The same procedure can be used to
clean the display window.
Caution: Clean the measurement tip with 90% or higher
Isopropyl Alcohol using a soft optical surface cleaning wipe.
WARNING: Do not immerse any part of the system in water
or other liquid. If liquids spill onto the unit, wipe with a damp
cloth and allow to dry prior to using the system.
WARNING: Do not attempt to clean and/or reuse the
BiliCal™ Individual Calibration Tip.
1014988
13.0 Maintenance and Repair and
Replacement
13.1
Hand Held Unit (HHU) Measurement Light Bulb
Replacement
If the instrument bulb is not working and error code E01 or
E10 (refer to the troubleshooting section 11.4) is displayed, it
is necessary to replace the measurement light bulb. A Re-
placement Bulb Kit (RI P/N B800-22) is available from
Respironics or your local distributor.
WARNING: The BiliChek
Analyzer is to be used only with the approved Replacement
Bulb Kit from Respironics. Do not over-tighten the lamp
cap.
Remove the battery pack from the bottom of the HHU.
Remove the protective label covering the lamp
assembly.
Remove the lamp cap using a flat tip screwdriver.
Grab the loop of the light bulb with forceps or similar
tool/instrument and gently pull to remove the lamp
assembly.
Insert the replacement lamp assembly using caution
not to touch bulb with fingers. Be sure to place bulb
into socket firmly. DO NOT use the lamp
cap to tighten bulb into place.
Apply the new lamp cap and protective label furnished
in the kit.
Insert a fully charged battery pack into the bottom of
the HHU.
BiliChek Service Manual
Non-Invasive Bilirubin
®
32