11.0 Troubleshooting Guide

11.1 Troubleshooting Guide
Display
Indication
Display Window
Window
E00
No user set-up information identified.
E01
Calibration measurement is unacceptable.
E02
Patient measurement is unacceptable.
E03
BiliCal Individual Calibration Tip removed
between calibration and measurement scan.
E04
Too much movement during measurement.
E05
High Measurement is out of clinically proven range.
E10
Light bulb is burned out.
E99
System error is detected.
Low battery pack. Maximum of 10 measurements left.
Discharged Battery pack.
No measurements can be taken.
1014988
Indication
BiliChek Service Manual
Solution
Solution
Install a charged battery pack into the HHU. Press
F1 and F2 at the same time, hold for one second and then release. This
places the HHU in the Set-up mode. Enter settings specific to your
institution. See section 10.0 of this manual for Set-up Mode
instructions.
Press and release F1 and F2 at the same time.
Ensure BiliCal tip is in place and retry the calibration. If the error
continues: 1. Place a fully charged Battery Pack in the HHU, install a
new BiliCal Tip, and try again to calibrate several times.
2. Remove and rotate the BiliCal tip and retry. 3. Replace the measurement
light bulb (see Section 14.1 for instructions) and retry. If the error is still
present, the HHU has malfunctioned and requires service. Call
Respironics or your distributor.
Press and release F1 and F2 at the same time. Take another
measurement. If E02 occurs again, replace the BiliCal tip, repeat
calibration and retry the measurement. If the error is still present, the
HHU has malfunctioned and requires service. Call
Respironics or your distributor.
Press and release F1 and F2 at the same time.
Remove the BiliCal Individual Calibration Tip.
Replace with a new tip and recalibrate.
Press and release F1 and F2 at the same time.
Perform another measurement.
Obtain a Total Serum Bilirubin level (TSB)
Replace measurement light bulb (see section 14.1 for instructions). If
error continues, the HHU has malfunctioned and requires service. Call
Respironics or your distributor.
Call for Service.
Replace current battery pack with a fully
charged battery pack from the charger base and place the discharged
battery in well “B” for charging.
Replace current battery pack with a fully charged one from the charger
base. Place the discharged battery pack into well “B” for charging.