AOIP CALYS 50 Calibration

On-site multifunction calibrator
Hide thumbs Also See for CALYS 50:

Advertisement

Quick Links

CALIBRATION
CALYS 50
On-site multifunction calibrator
Simultaneous Measurement and Generation
Protected for on-site use
Easy connection system
Designed with a close collaboration with industrials, Calys 50 integrates all the necessary
functions for adjustment and maintenance of process.
Its ergonomic design and its embedded software allow Calys 50 to be a high performance
calibrator and very easy to use.
IP 54, full protected by an anti-shock sheath, with "easy-connect®" terminals integrated, a
very contrasted backlight display, it will be comfortable to use in all conditions
Its elastomer keypad protects it from dirty and grease marks, the keys allow to use Calys 50
with gloves. It has 10 working configurations recorded by user or user group, to always find
the good configuration for designed and repetitive jobs.
Ed 02/10/04
Document susceptible d'être modifié

Advertisement

loading
Need help?

Need help?

Do you have a question about the CALYS 50 and is the answer not in the manual?

Questions and answers

Summary of Contents for AOIP CALYS 50

  • Page 1 Its elastomer keypad protects it from dirty and grease marks, the keys allow to use Calys 50 with gloves. It has 10 working configurations recorded by user or user group, to always find the good configuration for designed and repetitive jobs.
  • Page 2 Protected by a polycarbonat Power supply USB connector Calys 50 is delivered in standard with a strap and a stand for desktop use, a quick battery charger, and a set of 6 measuring cables with crocodile clips Ed 02/10/04 Document susceptible d'être modifié...
  • Page 3: Display Resolution

    Display resolution: Calys 50 allows the digit number after the dot to be selected: This function is justified by the needs of users who want or not to display the best resolution for calibration or on the contrary limit it for simple verifications.
  • Page 4 CALIBRATION Functions and performances:@23°C ±5°C Accuracies are given in % of reading(a CALYS 50 display) + fixed value Direct current: Measurement DC current: Emission Calys 50 is able to measure up to 50 mA, with or without power supply of loop (24V power supply).
  • Page 5 CALIBRATION Temperature Resistive probes: Measurement and emission Probe type Range Resolution Accuracy / 1yr Resolution Accuracy / 1yr Measurement Measurement Emission Emission Pt 50(α = 3850) - 220°C + 1 200°C 0,01°C 0,012 % L+ 0,06°C 0,03°C 0,014 % L+ 0,18°C - 220°C + 1 200°C 0,01°C 0,012 % L+ 0,05°C...
  • Page 6: Additional Functions

    Average, minimum, maximum and also number of measurements done are always displayed. Reset key allows values to be updated. Calys 50 is supplied in standard with 6 testing leads, a quick battery charging system, an instruction manual Instructions to order: On-site calibrator...