Summary of Contents for Mopec BF400

  • Page 1: User Manual

    USER MANUAL BF Series Down/Back Draft Workstations BF400, BF450, BF500, BF510, BF550 Rev 10-28-16 © www.mopec.com...
  • Page 2 P a g e...
  • Page 3: Table Of Contents

    TABLE OF CONTENTS UNPACKAGING YOUR PRODUCT..…………………………………………………4 INSTALLATION………..…………………………………………………………….……..5 PRODUCT DRAWINGS BF400 ROUGH IN……………………………………………………………..7 BF450 ROUGH IN……………………………………………………………..8 BF500 ROUGH IN (w Optional Leg Frame)..……………………..9 BF510 ROUGH IN…………………………………………………..………..10 BF550 ROUGH IN ………………………………………………..………….11 INTRODUCTION ……………………………………………………..…………………..12 PRODUCT DRAWINGS BF400 …………………………………………………….……………………..13 BF450 …………………………………………………….……………………..14 BF500 …………………………… ……….…………….………………………15 BF510 ……………………………………………………………………………16...
  • Page 4 Mopec and the freight carrier. Never discard the shipping container even if it is damaged beyond recognition. Have the delivery driver note any suspected damage on the Bill of Lading and sign it. Mopec will help assist in filing a claim for product repair and/or replacement.
  • Page 5: Installation

    The BF500, BF510 and BF550 workstations are equipped with 1.5” Diameter drain and connections. DO NOT modify the drain line or connections without contacting MOPEC first. The Units are built in accordance with the rough in drawings specific to the unit. The height is per the rough in drawings provided at the time of the order.
  • Page 6 Recirculation BF400, BF500 and BF510 the filters will need to be installed. You should have a minimum of 2 inches or more behind the workstation for the filtered air to discharge. Air flow is generated by multiple fans in the recirculation models. They are controlled by a rheostat to control volume/speed.
  • Page 7: Bf400

    Rough In Drawings BF400 P a g e...
  • Page 8: Bf450

    BF450 P a g e...
  • Page 9: Bf500

    BF500 P a g e...
  • Page 10: Bf510

    BF510 P a g e | 10...
  • Page 11: Bf550

    BF550 P a g e | 11...
  • Page 12 The BF400 and BF500 workstations are designed with a down and back draft ventilated work surface. The design allows air to flow through the grid plate. The BF400 is designed with a catch pan to collect fluids if necessary while the BF500 is freestanding with a built-in drain connection that fluids can flow directly into the drain.
  • Page 13 UNIT DRAWINGS BF400 P a g e | 13...
  • Page 14 BF450 P a g e | 14...
  • Page 15 BF500 P a g e | 15...
  • Page 16 BF510 P a g e | 16...
  • Page 17 BF550 P a g e | 17...
  • Page 18: Features

    Perforated Downdraft Grid Plate Circuit Breaker Switch Unique features by unit consist of the following: BF400 Fan System: Dual Exhaust Blowers, 300 CFM. Both blower’s velocity is adjustable speed, controlled by a rheostat. Filters: 7” W x 14” L x 1” (183 cm x 35 cm x 2.5 cm) Front Access (Fine Particle Activated Charcoal Media or Potassium Permanganate) Storage Shelf: Top shelf is 12”...
  • Page 19: Air Handling

    Therefore, assumptions are made regarding grossing practices and ultimately formaldehyde concentrations. Two (2) blowers are supplied with the BF400, BF500 and BF510 are designed with an adjustable control to regulate the speed and consequentially the airflow. The BF450 and BF550 workstations are in house ventilation and do not have...
  • Page 20: Cleaning And Maintenance

    CLEANING AND MAINTENANCE EVALUATING FILTERS FOR REPLACEMENT The filters in your MB unit contain alumina pellets impregnated with potassium permanganate, KMnO , which is a fast oxidizer. Formaldehyde passing through the filter is converted to carbon dioxide and water. The filter’s life depends entirely on the amount of formaldehyde fumes passing through the filter.
  • Page 21: Stainless Steel Care And Maintenance

    3) Rinse the surface after using disinfectant. In Pathology and other medical areas, the act of disinfecting is desired. There are several ways to do this including using Mopec’s Bench Spray & Wipe Disinfectant. Labs use any number of different products including 10% bleach or other disinfectant sprays and wipes. For the most part, each one of these has high salt contents and lower PH levels to aid with disinfection.
  • Page 22 Conclusions & Suggestions We are confident and can assure you that if you institute the suggestions detailed above that your Mopec Grossing station will look as it did the day it arrived. Do not assume it's the cleaner. If you do have some spotting or staining, and you've followed all the suggestions, it may not be the cleaner.
  • Page 23 deposits (DO NOT USE STEEL WOOL OR BRILLO PADS). If a more severe treatment is needed to mask coarse scratches or physical damage on a surface, use the finest abrasive medium consistent with covering the damage marks. With directional brushed and polished finishes, align and blend the new "scratch pattern"...
  • Page 24: Parts

    USER PARTS Replacement parts are available from Mopec They can be ordered by contacting Mopec at 800-362-8491. POTASSIUM PERMANGANATE FILTER BF007 Dimensions: 14" (35cm) L x 7" (17cm) W x 1" (2.54cm) H ACTIVATED CARBON FILTERS BF008 Dimensions: 14" (35cm) L x 7" (17cm) W x 1" (2.54cm) H...
  • Page 25 PREVENTITIVE MAINTENANCE CHECKS Procedure: 1. Visually check the exterior of equipment for any signs of damage. 2. Visually check the condition of the power cord and plug(s) for cracks, cuts, bare or broken wires and signs of excessive heat (discoloration). 3.
  • Page 26: Warranty

    MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. To the extent that Mopec is acting as a supplier of Products manufactured by a third party, the Products will be warranted only to the extent that they are warranted by their manufacturers and Buyer (or its customer) agrees to look solely to the Product manufacturer for all warranty claims.
  • Page 27: Safety Labels

    SAFETY LABELS BEFORE SERVICING THE UNIT LOOK FOR AND HEED THE FOLLOWING LABEL Symbols that may be found on the Equipment P a g e | 27...
  • Page 28: Trouble Shooting Guide

    TROUBLE SHOOTING Problem Possible Solution My Unit does not turn on Assure your facility circuit breaker has not been tripped. Assure the G.F.C.I. has not been tripped (off) – Reset to on. y hand spray does not work Assure the water valve from your facility is on. Assure the concealed shut off valves to the hand spray are on.
  • Page 29: Electrical Diagram

    ELECTRICAL DIAGRAM P a g e | 29...
  • Page 30: Certificates

    CERTIFICATES P a g e | 30...
  • Page 31 P a g e | 31...