SPECIFICATIONS, MODEL 5170 & 5180 GAUSS / TESLAMETER
DC MODE, Low Range:
Mid & High Ranges:
AC MODE (for sinewave >6G or 0.6mT) {1}
10 - 20 Hz
20 - 20,000 Hz :
20kHz to 25kHz (1x axial probe only)
{1} additional ±8 counts tolerance for signals
< 4% of lowest range and >2kHz
ACCURACY, Analog Output of Model 5180
DC MODE
Low Range:
Mid & High Ranges:
AC MODE : (for sinewave >6G or 0.6mT) {2}
Analog output -3dB point is approximately 22kHz
Delay Time: 30µs
Sample Rate: 8µs with variable moving average filtering
Risetime: DC Mode = 250µs
AC Mode and AC Peak Hold; from 30µs to 2ms depending on freq.
{2} additional ±8 mV tolerance for signals
< 4% of lowest range and >2kHz
Battery Type: 1.5V AA (x4)
Battery Life: 20 hours typical, alkaline batteries
Auxilliary Power: +5 Vdc, 300 mA (Regulated)
Auxilliary Power Connector: Standard 2.5mm I.D. / 5.5mm
O.D. connector. Center post is (+) polarity.
Analog Output Connector: BNC
Operating Temperature: 0 to +50ºC (+32 to +122ºF)
Storage Temperature: -25 to +70ºC (-13 to +158ºF)
Meter Dimensions: Length: 17.5 cm (6.9 in)
Width: 9.9 cm (3.9 in)
Height: 3.6 cm (1.4 in)
Weight:
Meter w/batteries: 400 g (14 oz.)
Shipping: 1.59 kg (3 lb., 8 oz.)
5170
5180
1.25 + 4
0.8 + 4
1.00 + 3
0.6 + 3
3.50 + 8
3.0 + 8
2.50 + 5
2.0 + 5
N/A
3.0 + 5
(±% of Reading ± mV)
0.75 + 6
0.50 + 5
10 - 20Hz
3.00 + 8
20 - 5000Hz
2.50 + 5
DC peak hold = 60µs
without probe, 23 ±3ºC, RH <85%
Probe accuracy must be added to meter accuracy to determine
overall accuracy
Warmup Time To Rated Accuracy:
Min / Max Hold Acquisition Time:
Peak Hold Acquisition Time:
Analog Output Scaling:
Analog Output Noise:
Analog Output Load:
Accuracy Change with Temperature (not including
probe)
0-20ºC and 26-50ºC (typical)
Low Range: ±0.25% + 3.0 counts / ºC
Mid & High Ranges: ±0.02% + 0.2 counts / ºC
Regulatory Information:
Compliance was demonstrated to the following specifications as listed in the
official Journal of the European Communities:
EN 61326-1:1997+A1:1998 Electrical equipment for measurement, control
and laboratory use EMC requirements
Immunity:
IEC 61000-4-2
Electrostatic Discharge (ESD)
IEC 61000-4-3
Electromagnetic Field (RF)
Emissions:
EN 55022 Class B Radiated and conducted emissions
Communications Port:
Format: Universal Serial Bus (USB)
Lines supported: Transmit, receive, common.
Connector type: Mini USB Type B
Cable length: 3 m (9.8 ft.) maximum
Standards supported: IEEE-1987.2, SCPI-1991
5 minutes
DC mode: 100 ms typical
AC mode: 700 ms at 10 Hz
to 500 ms at 300 Hz
250 ms at 300 - 4000 Hz
100 ms > 4kHz
DC or AC mode: 128 µs minimum
DC mode: ± 3 Vdc
AC mode: 3 Vrms
(Hi- Range) 1.5mV rms
(Mid- Range) 4.5mV rms
(Low- Range) 15.5mV rms
10 kOhm min, 100 pF max.
2-2