Philips PSD1211 User Manual page 151

Bluecontrol
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Problem
The Philips
BlueControl
suddenly switches
off with a long
beep during
treatment and the
LED indicator lights
up solid orange.
The Philips
BlueControl
suddenly
interrupts the
treatment and
slowly flashes blue.
Possible cause
The sensor is
blocked.
The rechargeable
battery is empty.
The temperature
of the Philips
BlueControl
itself is too high
and the Philips
BlueControl
switched off for
safety reasons.
The device
measured a
too high skin
temperature
and paused the
treatment to allow
the skin to cool
down.
ENGLISH
Solution
Check the position of the Philips
BlueControl in the fixation
strap. Make sure the material
of the fixation strap or clothing
does not cover the sensor. If the
sensor is not covered but the
Philips BlueControl still switches
off, the skin temperature is too
high. The Philips BlueControl will
switch on automatically when
the skin has cooled down.
The LED indicator flashes
red. Charge the Philips
BlueControl (see chapter
'Charging').
Let the Philips BlueControl
cool down for 10 minutes.
Put the Philips BlueControl
back onto your skin and press
the on/off button to resume
treatment.
No action required.
The treatment restarts
automatically after the device
measures a skin temperature
in the safe range.
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents