Subject/Attribute
Custom comments
Control lot verification
Control result display
Critical range display
Critical range HI limit
Critical range LO limit
Critical range message enabled
Critical range text
Normal range HI limit
Normal range LO limit
STAT tests allowed
Number of STAT tests allowed
Reportable range HI limit
Reportable range LO limit
Reportable range message
Reportable range message text
Strip limitations warning
Strip lot verification
150
Range
0: enabled
1: disabled
0: display only
1: yes / no confirmation
2: prompt for entry
3: scan only
0: value (numeric)
1: PASS / FAIL
0: value (numeric)
1: HI / LO
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10 – 600 mg/dL
0.6 – 33.3 mmol/L
Whether to display the
out-of-critical range
warning message (1) or
not (0)
0 – 100 characters
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10 – 600 mg/dL
0.6 – 33.3 mmol/L
0: no
1: yes
0 – 9
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10 – 600 mg/dL
0.6 – 33.3 mmol/L
Whether to display a
reportable range
message (1) or not (0)
0 – 100 characters
Configuration, whether to
display the strip limita-
tions warning (1) or not
(0).
0: display only
1: yes/no confirmation
2: list selection
3: scan only
Default
Device
0
N
1
Y
0
N
0
N
600 mg/dL
Y
33.3 mmol/L
10mg/dL
Y
0.6 mmol/L
1
N
""
N
600 mg/dL
Y
33.3 mmol/L
10mg/dL
Y
0.6 mmol/L
1
Y
9
Y
600 mg/dL
Y
33.3 mmol/L
10mg/dL
Y
0.6 mmol/L
1
N
""
N
0
Y
1
Y
DMS
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y

Hide quick links: