Analogue Output Option

Panel mounting

A panel mounted instrument must have the rubber feet removed for the side
plates to be secured. The instrument fits into a panel cut-out the side and
front plates secure the instrument to the panel. It is important that a panel
mount installation provides enough circulation of air to cool the instrument.
Procedure
To fit this option requires a panel cut-out of the dimensions shown in
1.
Remove the rubber inserts in the feet of the instrument.
2.
Unscrew and remove two 45mm screws (2) attaching two feet on one
side of the instrument casing. Retain the two 45mm screws for the
next step.
3.
Fit the side plate (1) to the side of the instrument casing and secure
with the two 45mm screws.
4.
Repeat steps 2 and 3 for the other side plate.
5.
Locate the assembled instrument behind the cut-out panel and align
the four 3.5mm holes in the panel and the holes in the flanges of the
side plates (1).
6.
Locate the front plate (3) over the front of the protruding instrument
and secure the front plate with the four screws (4).
and
2a
Key to
1 side plate
2 front plate
3 screw 3.5mm
Druck DPI 142 User Manual
3
2a
.
K0343 Issue No. 3