Procon Druck DPI 142 User Manual

Precision barometer

Advertisement

GE
Sensing & Inspection Technologies
Druck DPI 142
Precision Barometer
User Manual - K0343
FS
110 0 mb ar g
H
T
1 0 0 3 . 6 9 6
s e tt in g s
110 0 mb ar abs
4 2
www.proconsystems.com

Advertisement

Table of Contents
loading

Summary of Contents for Procon Druck DPI 142

  • Page 1 Sensing & Inspection Technologies Druck DPI 142 Precision Barometer User Manual - K0343 110 0 mb ar abs 110 0 mb ar g 1 0 0 3 . 6 9 6 s e tt in g s www.proconsystems.com...
  • Page 2: Approved Service Agents

    Approved Service Agents www.gesensinginspection.com © The General Electric Company. All rights reserved...
  • Page 5 Druck DPI 142 User Manual General Introduction This manual provides operating instructions for the DPI 142 Precision Barometer compatible with the requirements of operating the instrument. Safety The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. The user must not use this equipment for any other purpose than that stated.
  • Page 6: Table Of Contents

    Druck DPI 142 User Manual ABBREVIATIONS The following abbreviations are used in this publication. Note: Abbreviations are the same in the singular and plural. absolute maximum working pressure °C degrees Celsius mbar millibar digital pressure indicator (GE product) mm millimetre...
  • Page 7: Introduction

    Druck DPI 142 User Manual Introduction The Druck DPI 142 precision, barometric pressure indicator uses a silicon resonant pressure transducer to produce pressure readings in units of pressure measurement and aeronautical units. The instrument is contained in a moulded plastic case with integral rubber feet for workbench surface use.
  • Page 8: Installation

    Druck DPI 142 User Manual Analogue Option Electrical Specification Isolated Voltage Output:- Output Impedance............... 5 Ohm Maximum Load Capacitance............10 nF Isolated Current Output:- Maximum Load Impedance............. 900 Ohm Maximum Load Capacitance............10 nF Accuracy (including pressure measurement uncertainty) All voltage and current ranges ..±0.05 % FS (18° to 28°C, 12 months) Update rate ............
  • Page 9: Analogue Output Option

    Druck DPI 142 User Manual Analogue Output Option Panel mounting A panel mounted instrument must have the rubber feet removed for the side plates to be secured. The instrument fits into a panel cut-out the side and front plates secure the instrument to the panel. It is important that a panel mount installation provides enough circulation of air to cool the instrument.
  • Page 10: Bench Stand

    Druck DPI 142 User Manual Bench stand A bench stand lifts the front of the instrument providing a better angle of display and key-pad access for the user. Key to 1 stand 2 screw 45 mm 3 foot 4 clip (left and right) 5 insert, rubber K0343 Issue No.
  • Page 11: Operation

    Druck DPI 142 User Manual Operation Key to 1 esc key 6 soft key right 2 menu OK key 7 soft key left 3 zero key 8 display 4 hold key 9 up or increase key 5 down or decrease key...
  • Page 12: Barometer

    Druck DPI 142 User Manual Barometer » 1100 Barometer 1003.696 mbar a QFE settings mbar units filter barometer mode mmH O mH O inHg inH O menu torr lbft² inHg kgcm² kgm² mmHg cmHg user1 menu user2 units menu menu...
  • Page 13: Barometer Settings

    Druck DPI 142 User Manual Barometer settings °C menu °F menu settings +020.0 temp. (T): °C –00100.0 height (h): +020.0 temp.: °C edit units –00100.0 height : menu °C °F menu settings +020.0 temp. (T): °C +00150.0 height (H): +020.0 temp.:...
  • Page 14: Height Reference

    Druck DPI 142 User Manual Height reference barometer installed in control tower. height from the barometer installation to the reference point (for example: the runway). Note: h is negative if the reference point is below the barometer installation. height of the barometer installation above or below standard sea level datum - mean sea level.
  • Page 15: User Set-Up

    Druck DPI 142 User Manual User Set-up » 1100 Barometer barometer user menu barograph 1003.696 supervisor altitude mbar a QFE calibration pressure leak test settings status altitude leak test menu set-up status menu display contrast display backlight menu resolution software build...
  • Page 16: Barograph

    Druck DPI 142 User Manual Barograph » 0.005 full scale altitude 992.837 airspeed menu menu mach -0.005 cur. 0.99385 settings settings barograph mean : 0.99000 units time settings mbar mmH O inHg inH O inH O4 ftH O torr ftH O4 lbft²...
  • Page 17: Altitude

    Druck DPI 142 User Manual Altitude » 105000 ft full scale set-up 549.152 16000 indicator menu menu leak test units settings altitude feet metres menu K0343 Issue No. 3...
  • Page 18: Pressure Leak Test

    Druck DPI 142 User Manual Pressure Leak Test barometer barograph » 1100 Barometer 1100 Leak Test altitude 1003.696 menu 1003.696 pressure leak test menu test time (s): mbar a mbar a QFE altitude leak test settings settings start set-up 1100 Leak Test 1003.696...
  • Page 19: Supervisor Set-Up

    Druck DPI 142 User Manual Supervisor Set-up K0343 Issue No. 3...
  • Page 20: Communication Option

    Druck DPI 142 User Manual Supervisor Set-up (continued) Communications Option K0343 Issue No. 3...
  • Page 21: Analogue Output Option

    Druck DPI 142 User Manual Supervisor Set-up (continued) Analogue Output Option K0343 Issue No. 3...
  • Page 22 Druck DPI 142 User Manual intentionally left blank K0343 Issue No. 3...