Installation; Packaging List; Connections - Procon Druck DPI 142 User Manual

Precision barometer
Table of Contents

Advertisement

Druck DPI 142 User Manual
Analogue Option Electrical Specification
Isolated Voltage Output:-
Output Impedance....................................................................... 5 Ohm
Maximum Load Capacitance.........................................................10 nF
Isolated Current Output:-
Maximum Load Impedance..................................................... 900 Ohm
Maximum Load Capacitance.........................................................10 nF
Accuracy
(including pressure measurement uncertainty)
All voltage and current ranges ....±0.05 % FS (18° to 28°C, 12 months)
Update rate ...................................................... 30 updates per second

Installation

1 power supply
2 external sensor connector 5 G1/8 or 1/8 NPT pressure connector
3 IEEE 488
Packaging List - DPI 142
i)
DPI 142 Precision, Barometric Pressure Indicator.
ii)
Power supply 191-363.
iii)
User manual (this document).
iv)
Pressure indicator calibration certificate.
vi)
Cable, RS232.

Connections

Note:
Switch off the power supply before connecting or disconnecting the
instrument. Isolate the pneumatic pressures and depressurise the
pipes before connecting or disconnecting the instrument.
1.
Use an appropriate sealing method for all pressure connections.
2.
Before use, make sure the SELV power adaptor supplied with the
instrument is correct for the power supply voltage. The Safety Extra
Low Voltage (SELV) power adaptor complies with EN61010 (including
safety requirements for laboratory instruments).
Note: The instrument can be powered from other AC or DC power supplies
in the range 11 to 26 Volts. It is the user's responsibility to make sure
the power supply is safe.
3.
Connect the power adaptor to the instrument and switch the power
supply on.
K0343 Issue No. 3
Key to
4 RS232 connector
6 analogue output connector
2

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Procon Druck DPI 142