2
Operation
To measure saturation
To measure temperature
To view other parameters while measuring
42
1 When the DO probe is used for the first time, rinse it with
distilled water and put it in sample solution.
2 Calibrate the meter. See
3 From the initial state display, press [Setup] and select a
measuring mode and the SATU parameter. See
Mode"
on page 26.
4 Put the probe in the sample solution and press [Measure] to
start taking a measurement according to the selected mode.
5 After a measurement, press [Save] to save data. Press [Output]
to print data. Press [Cancel] to end measurement.
In any measuring state, the meter directly displays current
solution temperature.
During a measurement, the user still can view other parameter
not currently in use without changing the operational mode. For
instance, while measuring with the DO parameter, a user can
view, save, and print the SATU parameter.
1 While performing a measurement, press [Mode]. The screen
display highlights the measuring window as shown in
Figure
24.
Figure 24
Viewing parameters while measuring
"Calibrate the DO Probe"
3200D Dissolved Oxygen Meter Operating Guide
on page 39.
"Measuring