Installation And Start-Up - LabRepCo GM-ULT-1786 Owner's Manual

Ultra low temperature freezer
Table of Contents

Advertisement

Power: The ULT requires a dedicated power outlet and circuit breaker: 20-amp circuit for
115V models and 15-amp circuit for 230V models.
Installation and start–up
Placement clearance: The freezer must have a clearance of 5" in the rear and 6"
on each side so that sufficient air circulation is maintained for the removal of
exchange heat.
Start –up: Plug the ULT into power and engage the power switch/breaker located above the
power cord. The freezer will start, and the control panel will illuminate in the default status with
the words SYSTEM NORMAL in the message line, which is the top line of the control display.
The other lines will display available operating modes. You will note a pointer (
line.
The third line is always the active line, which means that you will enter that mode or menu by
pressing ENTER.
The default status:
Temperature Set Point (SP)
High temp alarm
Low temp alarm
Password security
Access code
Part number 21729, Rev 4
READ THIS MANUAL
It is necessary to read and understand this instruction
manual because failure to do so may result in (1)
substandard performance of your freezer, (2) damage to
your freezer and stored items, and also (3) possible
injury to operating personnel.
NOTICE!
-80ᵒC
SP+10ᵒC
SP-10ᵒC
OFF
1234
3
˃
) on the third