T . O . 33K3-4-2590-1
4.1.10 The Meter Calibrator output readout must indicate 0.396 to 0.404.
4.1.11 Set the Meter Calibrator to ST DBY and disconnect from TI.
4.1.12 Return TI CH 1 VOLTSI DIV VAR control to the CAL (detent) positio n .
5

Hide quick links:

Advertisement