Biotek PowerWave Operator's Manual page 162

Microplate scanning spectrophotometer
Hide thumbs Also See for PowerWave:
Table of Contents

Advertisement

iii
NOTE : In cases where a sensor is not functioning, the motor will drive
Displayed Error
ERROR 0300
ERROR 0301
Errors 300 and 301 indicate that a particular axis has moved to a point
where the light beam from the optics is no longer detectable by the
measurement electronics. This error is usually only seen during the Auto
Cal sequence.
C-2
P
:
ROBABLE CAUSE
Carrier Axis Case - Bad sensor. The X-axis movement is limited
so that the optical sensor cannot be interrupted.
Filter Wheel Case - Bad sensor. The filter wheel movement is
limited so that the sensor cannot be interrupted. Or, the filter
wheel is not installed.
the axis to its mechanical stop and generate substantial noise.
P
:
ROBABLE CAUSE
Carrier - A loose belt, loose motor pulley or bad motor drive may
causes the carrier to ignore movement instructions.
Filter Wheel Motor - The filter wheel drive gear is loose or
motor drive failure is impeding filter wheel movement.
Potential Cause
Carrier failed to find light beam
Filter wheel did not find home
Error Codes