Page 1
CESI 8000 Plus High Performance Separation-ESI Module Preinstallation Manual RUO-IDV-05-3345-B | B11953AE August 2016...
Page 2
SCIEX warranties are limited to those express warranties provided at the time of sale or license of its products and are SCIEX s sole and exclusive representations, warranties, and obligations.
8000 Plus system. Note: Refer to the Safety Manual for instructions for safe use of the system. Note: If the site preparation tasks are not complete when the SCIEX Field Service Employee (FSE) arrives, then the scheduled installation will be postponed.
Technical Support SCIEX and its representatives maintain a staff of fully-trained service and technical specialists located throughout the world. They can answer questions about the system or any technical issues that might arise. For more information, visit the SCIEX Web site at sciex.com.
There is adequate lab space to accommodate the equipment. If the mass spectrometer inlet is lower or higher than the range of the CESI 8000 Plus cart, a table or stand of the appropriate height for the CESI 8000 Plus system is available.
Environmental Requirements on page Requirement Complete The altitude does not exceed 2000 m (6562 ft) above sea level. Temperature and humidity requirements have been met. CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 8 of 24 RUO-IDV-05-3345-B | B11953AE...
Any required equipment for the ion source or mass spectrometer is available at the site. Solutions and Equipment Requirements Refer to Customer-Supplied Solutions and Equipment on page Requirement Complete All of the required solutions and bottles are available. Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE 9 of 24...
Years of experience with mass spectrometry Years of experience with capillary electrophoresis CESI applications to be performed Will the CESI 8000 Plus system be used stand-alone? If the system will be used stand-alone, which detectors will be used? CESI 8000 Plus High Performance...
Completion date (yyyy-mm-dd) I acknowledge that all of the installation requirements, as specified in this document, have been met. FSE name Return date (yyyy-mm-dd) FSE e-mail Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE 11 of 24...
• The location is away from vibrating equipment, such as refrigerators or centrifuges. • For configurations where the CESI 8000 Plus controller is connected to the mass spectrometer acquisition computer with a 4.3 m (14 feet) cable, the maximum distance between the controller and the acquisition computer is 4 m (13 feet).
Height Requirements The alignment mark on the outside of the CESI 8000 Plus system is 36 cm (14 inches) from the bottom of the system. This mark must be vertically aligned with the mass spectrometer inlet. To align the two systems, the height of the cart provided with the CESI 8000 Plus system is adjustable.
Surge Suppression The CESI 8000 Plus system must be plugged into a high capacity electrical surge suppressor, to protect the system from voltage spikes and power surges. The maximum current is the sum of the maximum current ratings for all components.
Site Requirements Table A-1 CESI 8000 Plus Module Electrical Specifications Specification Value Nominal input voltage Instrument: 100 VAC to 240 VAC Cart: 100 VAC to 240 VAC Frequency 50 Hz or 60 Hz Maximum input current Instrument: 5 A Power Consumption Supply voltage must not exceed 10% of nominal.
• Relative humidity from 20% to 60%, non-condensing BioSafety Requirements The site must not be designated BioSafety Level 3 (BSL-3) or BioSafety Level 4 (BSL-4). SCIEX does not install, service, or repair SCIEX systems in areas designated BSL-3 or BSL-4.
The OptiMS Bruker MS Adapter for Bruker Mass Spectrometers (PN B86099) requires the NanoElectrospray capillary cap set (PN 212915), purchased from Bruker. For Bruker mass spectrometers that are close to the ground, the CESI 8000 Plus system requires a low table or stand (refer to Height Requirements on page 13).
• Glacial acetic acid (Sigma PN A6283 or equivalent) • Methanol (Fisher PN A454 or equivalent) • Double-deionized (DDI) water (MS grade water filtered through 0.2 μm filter and with resistance above 18 megaohms) CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 18 of 24...
Note: Environments with a Pollution Degree 2 rating include laboratories and sales and commercial areas. For more information, refer to the International Electrotechnical Commission standards IEC 61010-1 and IEC 60364. Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE...
91.4 cm × 73.7 cm (36 inches × 29 inches) Height Adjustable from 68.6 cm to 111.8 cm (27 inches to 44 inches) Weight 69.0 kg (152 pounds) CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 20 of 24 RUO-IDV-05-3345-B | B11953AE...
120 VAC at 60 Hz, optional 230 VAC at 50 Hz Supports 136.0 kg (300 pounds) Controller Specifications The CESI 8000 Plus High Performance Separation-ESI Module includes a controller, with the specifications below. Note: Specifications are subject to change without notice. Lenovo PC with 3.1 GHz processor Memory...
256 element diode array Detector wavelength accuracy 2 nm Detector wavelength range 190 nm to 600 nm Scan collection frequency 0.5 Hz to 32 Hz CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 22 of 24 RUO-IDV-05-3345-B | B11953AE...
Added required equipment for Bruker and Waters mass August 2016 B11953AE spectrometers. Added requirements for mass spectrometer performance test. Added requirements for system height. Added ventilation and waste collection requirements. CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 24 of 24 RUO-IDV-05-3345-B | B11953AE...