SCIEX CESI 8000 Plus Preinstallation Manual

SCIEX CESI 8000 Plus Preinstallation Manual

High performance separation-esi module
Hide thumbs Also See for CESI 8000 Plus:
Table of Contents

Advertisement

Quick Links

CESI 8000 Plus High Performance
Separation-ESI Module
Preinstallation Manual
RUO-IDV-05-3345-B | B11953AE
August 2016

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CESI 8000 Plus and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SCIEX CESI 8000 Plus

  • Page 1 CESI 8000 Plus High Performance Separation-ESI Module Preinstallation Manual RUO-IDV-05-3345-B | B11953AE August 2016...
  • Page 2 SCIEX warranties are limited to those express warranties provided at the time of sale or license of its products and are SCIEX s sole and exclusive representations, warranties, and obligations.
  • Page 3: Table Of Contents

    Surge Suppression..............................14 System Electrical Specifications..........................14 Ventilation and Waste Collection Requirements......................15 Ventilation Requirements............................15 Waste Collection Requirements..........................15 Environmental Requirements............................15 BioSafety Requirements.............................16 Customer Preparation Requirements..........................16 Mass Spectrometer Requirements............................16 Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE 3 of 24...
  • Page 4 Cart Specifications................................20 Controller Specifications..............................21 Detector Specifications..............................21 (Optional) UV Detector Specifications........................21 (Optional) Laser Induced Fluorescence (LIF) Detector Specifications................22 (Optional) Photo Diode Array (PDA) Detector Specifications..................22 Revision History............................24 CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 4 of 24 RUO-IDV-05-3345-B | B11953AE...
  • Page 5: Chapter 1 Introduction

    8000 Plus system. Note: Refer to the Safety Manual for instructions for safe use of the system. Note: If the site preparation tasks are not complete when the SCIEX Field Service Employee (FSE) arrives, then the scheduled installation will be postponed.
  • Page 6: Fse Responsibilities

    Technical Support SCIEX and its representatives maintain a staff of fully-trained service and technical specialists located throughout the world. They can answer questions about the system or any technical issues that might arise. For more information, visit the SCIEX Web site at sciex.com.
  • Page 7: Chapter 2 Site Planning Checklist

    There is adequate lab space to accommodate the equipment. If the mass spectrometer inlet is lower or higher than the range of the CESI 8000 Plus cart, a table or stand of the appropriate height for the CESI 8000 Plus system is available.
  • Page 8: Electrical Requirements

    Environmental Requirements on page Requirement Complete The altitude does not exceed 2000 m (6562 ft) above sea level. Temperature and humidity requirements have been met. CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 8 of 24 RUO-IDV-05-3345-B | B11953AE...
  • Page 9: Biosafety Requirements

    Any required equipment for the ion source or mass spectrometer is available at the site. Solutions and Equipment Requirements Refer to Customer-Supplied Solutions and Equipment on page Requirement Complete All of the required solutions and bottles are available. Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE 9 of 24...
  • Page 10: Mass Spectrometer Information

    Years of experience with mass spectrometry Years of experience with capillary electrophoresis CESI applications to be performed Will the CESI 8000 Plus system be used stand-alone? If the system will be used stand-alone, which detectors will be used? CESI 8000 Plus High Performance...
  • Page 11: Comments And Exceptions

    Completion date (yyyy-mm-dd) I acknowledge that all of the installation requirements, as specified in this document, have been met. FSE name Return date (yyyy-mm-dd) FSE e-mail Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE 11 of 24...
  • Page 12: Appendix A Site Requirements

    • The location is away from vibrating equipment, such as refrigerators or centrifuges. • For configurations where the CESI 8000 Plus controller is connected to the mass spectrometer acquisition computer with a 4.3 m (14 feet) cable, the maximum distance between the controller and the acquisition computer is 4 m (13 feet).
  • Page 13: Height Requirements

    Height Requirements The alignment mark on the outside of the CESI 8000 Plus system is 36 cm (14 inches) from the bottom of the system. This mark must be vertically aligned with the mass spectrometer inlet. To align the two systems, the height of the cart provided with the CESI 8000 Plus system is adjustable.
  • Page 14: Electrical Requirements

    Surge Suppression The CESI 8000 Plus system must be plugged into a high capacity electrical surge suppressor, to protect the system from voltage spikes and power surges. The maximum current is the sum of the maximum current ratings for all components.
  • Page 15: Ventilation And Waste Collection Requirements

    Site Requirements Table A-1 CESI 8000 Plus Module Electrical Specifications Specification Value Nominal input voltage Instrument: 100 VAC to 240 VAC Cart: 100 VAC to 240 VAC Frequency 50 Hz or 60 Hz Maximum input current Instrument: 5 A Power Consumption Supply voltage must not exceed 10% of nominal.
  • Page 16: Biosafety Requirements

    • Relative humidity from 20% to 60%, non-condensing BioSafety Requirements The site must not be designated BioSafety Level 3 (BSL-3) or BioSafety Level 4 (BSL-4). SCIEX does not install, service, or repair SCIEX systems in areas designated BSL-3 or BSL-4.
  • Page 17: Thermo Scientific Mass Spectrometers

    The OptiMS Bruker MS Adapter for Bruker Mass Spectrometers (PN B86099) requires the NanoElectrospray capillary cap set (PN 212915), purchased from Bruker. For Bruker mass spectrometers that are close to the ground, the CESI 8000 Plus system requires a low table or stand (refer to Height Requirements on page 13).
  • Page 18: Separation-Esi Module

    • Glacial acetic acid (Sigma PN A6283 or equivalent) • Methanol (Fisher PN A454 or equivalent) • Double-deionized (DDI) water (MS grade water filtered through 0.2 μm filter and with resistance above 18 megaohms) CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 18 of 24...
  • Page 19: Appendix B Equipment Safety Categories

    Note: Environments with a Pollution Degree 2 rating include laboratories and sales and commercial areas. For more information, refer to the International Electrotechnical Commission standards IEC 61010-1 and IEC 60364. Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE...
  • Page 20: Appendix C System Specifications

    91.4 cm × 73.7 cm (36 inches × 29 inches) Height Adjustable from 68.6 cm to 111.8 cm (27 inches to 44 inches) Weight 69.0 kg (152 pounds) CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 20 of 24 RUO-IDV-05-3345-B | B11953AE...
  • Page 21: Controller Specifications

    120 VAC at 60 Hz, optional 230 VAC at 50 Hz Supports 136.0 kg (300 pounds) Controller Specifications The CESI 8000 Plus High Performance Separation-ESI Module includes a controller, with the specifications below. Note: Specifications are subject to change without notice. Lenovo PC with 3.1 GHz processor Memory...
  • Page 22: (Optional) Laser Induced Fluorescence (Lif) Detector Specifications

    256 element diode array Detector wavelength accuracy 2 nm Detector wavelength range 190 nm to 600 nm Scan collection frequency 0.5 Hz to 32 Hz CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 22 of 24 RUO-IDV-05-3345-B | B11953AE...
  • Page 23: Separation-Esi Module

    System Specifications UV source 30 W pre-aligned deuterium lamp UV source lifetime 1000 hours Preinstallation Manual CESI 8000 Plus High Performance Separation-ESI Module RUO-IDV-05-3345-B | B11953AE 23 of 24...
  • Page 24: Ruo-Idv-05-3345-B | B11953Ae

    Added required equipment for Bruker and Waters mass August 2016 B11953AE spectrometers. Added requirements for mass spectrometer performance test. Added requirements for system height. Added ventilation and waste collection requirements. CESI 8000 Plus High Performance Preinstallation Manual Separation-ESI Module 24 of 24 RUO-IDV-05-3345-B | B11953AE...