GlucoRx HCT Connect Owner's Manual page 21

Integrated multiparameter diabetes monitoring system
Hide thumbs Also See for HCT Connect:
Table of Contents

Advertisement

3. Obtaining a blood sample
Use your pre-set lancing device to puncture the desired site.
Wipe off the first drop of blood that appears with a clean
cotton swab. Gently squeeze the punctured area to obtain
another drop of blood. The volume of blood sample must
be at least 1.0 microliter (μL). Be careful NOT to smear the
blood sample.
4. Apply the Sample
Gently apply a drop of blood to the absorbent hole of your
test strip at a tilted angle. The confirmation window should be
completely filled if enough blood sample has been applied.
Wait until the "processing" message appears on the screen.
5. Read Your Result
The result of your test will appear after the device shows
the "Processing" message. This result will automatically be
stored in the memory.
6. Pull out the used test strip
Pull out the used test strip and dispose carefully in a sharps
bin.
21