Troubleshooting - Fluke 8021B Instruction Manual

Hide thumbs Also See for 8021B:
Table of Contents

Advertisement

8021
4-39.
Use
the
following procedure
to calibrate
the
802
IB:
NOTE
This
procedure assumes an ambient temperature of 23
±2°
C
(70 to 77°
F)
and
a
relative
humidity of
less
than
80%.
The
temperature of
the unit
should
be allowed
to stabilize
for
at least
30 minutes
before
calibration begins.
1.
Remove
the
top
cover
from
the
8021
B
using the access procedure given
earlier
in
this
section.
2.
Set
the
8021
B
power
switch
to
ON
and
select
the
200
mV DC
range.
3.
Set
the
output of
the
DC
Calibrator to
-H
90.0
mV
and
connect
it
to
the
802 IB
input terminals;
+
to V/ft,
and
-
to
COMMON.
4.
Adjust
the
DC CAL
pot
(R5), as
shown
in
Figure
4-1
,
fora
display
of
190.0 or
190.1.
(Use
a plastic
adjustment
tool
ora
plastic
screw
1
driver
for
all
adjustments.)
5.
Disconnect
the
DC
Calibrator
from
the
802 IB
input
terminals.
6.
Select the
200
mV
AC
range
on
the
802
1
B.
7.
Set the
output of
the
AC
Calibrator to 190
mV
at
100 Hz,
and
connect
it
to the
802IB
input terminals; V/fi
and
COMMON.
8.
Adjust
the
AC
CAL
pot
(R9)for
a
display of 190.0 (an occasional flash of
±1
digit
is
acceptable).
9.
Execute
the
performance
test
given
earlier to
ensure
that
all
fixed
range
resistors
and
other non-adjustable
components
are
operating within
their
specified
limits.
4-40.
TROUBLESHOOTING
CAUTION
Static
discharge
can
damage
MOS
components
contained
in
the
8021
B.
4-41.
When
troubleshooting
or repairing the
8021
B
use the precautions
listed
on
the
Static
Awareness
sheet to prevent
damage from
static
discharge.
Never remove,
install
or
otherwise connect or disconnect
components
without
first
setting the
802
1
B
power
switch
to
OFF.
4-42.
A
troubleshooting guide
for the
8021
B
is
given
in
Table
4-6.
To
properly use
the
guide complete
the
performance
test
given
earlier
in
this
section
and note any
discrepancies.
Then
locate
the
heading of
the
procedure
in
question
in
the
Test
and
Symptom
column
(Table
4-6).
Under
that
heading
isolate the
symptom
that
approximates
the
observed malfunction.
Possible causes are
listed
to the right
of
the selected
symptom.
Details necessary to
isolate a particular
cause
can be
derived
from
the
theory of
operation
in
Section
3
and
the
schematic diagram
in
Section
7.
4-10

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents