5. Position the aid on the 'T' on the tele-test handset. BTE aids should lie as flat as possible on the tele-test
surface, while custom aids should be oriented with their faceplate parallel to the top surface (if possible) and
touching it as shown.
6. Press
to start the test.
7. While maintaining the orientation of step 5, move the hearing instrument around on the tele-test handset
surface to maximize the number shown in the [Maximize Average-SPLITS Value] box.
8. Without moving the hearing instrument, press
SPLITS curve.
If an RTG has been previously obtained, the RSETS will be shown.

ANSI test loop test

ANSI S3.22 requires that the gain control be set to RTS before running this test. The reference test gain (RTG)
is required in order to calculate the RTLS. The easiest way to meet both requirements is to run a [ANSI] or
[ANSI / IEC] test before running this test. See ANSI 2003 and ANSI 2009 tests
1. First, complete an [ANSI] or [ANSI / IEC] test.
2. Click the right mouse button to raise the home menu, then select [Telecoil].
3. Set the aid to "T" (telecoil input).
4. Click [Start test loop test]. Setup instructions will be displayed.
5. Position the hearing instrument on the cross on the test box floor vertically oriented as worn by a seated
individual as shown.
6. Without moving the hearing instrument, press
curve or press
If an RTG has been previously obtained, the RTLS will be shown.
Verifit
®
User's Guide Version 4.8 © November 2016
to abort the test.
to store the Average HFA SPLITS value and run the
to run the SPL in a vertical magnetic field (SPLIV)