Instructions For Calibration - MedeScan RC008 User Manual

Hide thumbs Also See for RC008:
Table of Contents

Advertisement

Available languages

Available languages

Button MEM
Press the MEM button, the display shows the last
measured value from newest to oldest (memory
thermometer in the last 32 measurements).
Pressing MEM displays each temperature
thermometer from memory.
Button LANGUAGE
Press LANGUAGE to enable or disable the audio
guide thermometer.
NOTICE!
To obtain the internal temperature of the
body, always use the „BODY" mode. Please,
make sure to select „Sufrace temp" mode
for an external area reading.
CALIBRATION OF THE THERMOMETER
Thermometer factory-set. Depending on various
skin types and thickness, there may be temperature
differences.
Press the MODE and °C/°F buttons at the same time
for 3 seconds. The screen will then display „F-4".
Press MODE button to increase 0,1 °C, press °C/°F
button to reduce 0,1 °C.
Press MEM button to save the setting.
In the cases of seasonal or environmental changes a
verification and adjustment should be carried out.
38