Chapter 8 Glossary - Mantracourt ICA3 User Manual

Strain gauge or load cell embedded analogue amplifier
Hide thumbs Also See for ICA3:
Table of Contents

Advertisement

Chapter 8 Glossary

AWG
American Wire Gauge.
Background Noise
The total noise floor from all sources of interference in a measurement system,
independent of the presence of a data signal. (See Noise)
Bipolar
The ability of a signal conditioner to display both positive and negative
readings.
Bridge Resistance
The resistance measured across the excitation terminals of a Strain Gauge.
Calibration
Adjustment of an instrument or compiling a deviation chart so that it's reading
can be correlated to the actual value being measured.
CMR
The ability of an instrument to eliminate the effect of AC or DC noise between
(Common-Mode
signal and ground. Normally expressed in dB at DC to 60 Hz. One type of CMR is
Rejection)
specified between SIG LO and PWR GND. In differential meters, a second type
of CMR is specified between SIG LO and ANA GND (METER GND).
Common Mode
The ability of an instrument to reject interference from a common voltage at
Rejection Ratio
its input terminals with relation to ground. Usually expressed in db (decibels).
Drift
Change of a reading/set point value over periods due to several factors
including change in ambient temperature, time and line voltage.
Excitation
The external application of electrical voltage current applied to a transducer
for normal operation.
Fine Adjustment
Zero and Span calibration have a Fine Adjustment to give accuracy to the
calibration. These are potentiometers P1 and P2 for Span and Zero
respectively.
Full Bridge
A Wheatstone Bridge configuration utilizing four active elements or Strain
Gauges.
Full Range Output
The algebraic difference between the minimum output and maximum output.
Gain
Gain is otherwise identified as SPAN. It relates to the proportional output to
the sensor input. Calibration of the ICA is determined by setting the Gain
(Span) and Offset (Zero). The amount of amplification used in an electrical
circuit.
Ground
The electrical neutral line having the same potential as the surrounding
ground.
Linearity
The closeness of a calibration curve to a specified straight line. Linearity is
expressed as the maximum deviation of any calibration point on a specified
straight line during any one calibration cycle.
Load
The electrical demand of a process expressed as power (watts), current (amps)
or resistance (ohms).
Load Impedance
The impedance presented to the output terminals of a transducer by the
associated external circuitry.
Load Cell
The load cell is one of a series of STRAIN GAUGE sensors that the ICA input is
designed to accept. (Torque Sensor, Pressure & temperature transducers).
Millivolt
One thousandth of a volt, 10
Noise
An unwanted electrical interference on the signal wires.
Null
A condition, such as balance, which results in a minimum absolute value of
output.
Offset
Offset is otherwise identified as ZERO. Calibration of the ICA is determined by
setting the Offset (Zero) and Gain (Span).
Potentiometer
Two potentiometers (variable resistors) are used in the ICA for fine calibration.
Pressure
The Pressure Transducer is one of a series of Strain Gauge sensors that the ICA
Transducer
input is designed to accept. (Torque Sensor, Load Cell and Temperature
Transducers).
Proportional
The Voltage or Current outputs are calibrated to be directly proportional to the
Outputs
input from the sensor. The output is, within the sensor limits, taken as linear
and no linearity compensation is required within the ICA.
Mantracourt Electronics Limited ICA3 & 6 S and H Versions User Manual
-3
volts symbol mV.
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ica6sIca6h

Table of Contents