Calibration; Print Setups - Accurist W3300 Series Operating Manual

Table of Contents

Advertisement

4. CALIBRATION

The following calibration procedure can be used to calibrate the balance to a
known calibration weight standard. Before proceeding, the balance must be
located in a stable environment, equilibrated to the ambient temperature.
The leveling bubble indicator should show that the balance is level and the
weighing pan should be empty and clean.
1. Press the CAL key, then press TARE to scroll through available
calibration weight options (for example 500, 200, 100).
2. When a suitable calibration weight is displayed, wait for "OK" to be
displayed. This indicates that the balance is stable.
3. Press "CAL" again, the display will show "ACAL" to indicate that balance
is ready for calibration.
4. Place the appropriate calibration weight onto the weighing pan and
wait for "OK" to display.
5. The calibrated value (for example 100.00 g) will be displayed when the
calibration is complete.
6. Remove the weight, and the display should return to 0.00.
7. The balance is calibrated.

5. Print Setups

There are three print types:
STABLE Print: print a stable reading when it is attained.
INSTANT Print: print immediately after press the Printkey.
INTERVAL Print: print at predetermined time intervals.
The number of line feeds also is set for label printing.
Note:
The print function is separate from the line feed setup, i.e, set the print function first
then set the number of line feeds.
To set the print type, use the following procedure:
1. Press and hold PRINT key and also press the % PERCENT key, the balance
reads "PrlnT"
9

Hide quick links:

Advertisement

Table of Contents
loading