·Unstable power supply
·The object weighed is
·The balance is not
Weighing
value is not
calibrated
·The display is not tared
accurate
before weighing
·The balance is not
properly leveled

8. CARE AND MAINTENANCE

The Accururis Analytical Balance is a high precision instrument that
must be handled carefully and properly maintained.
·Do not use sharp objects(such as a pen or pencil) on the keypad.
·Do not let objects fall on the weighing pan, otherwise the weighing
system can be damaged;
·Do not expose the balance to high temperatures or dust;
·Do not disassemble the balance without proper training or
instructions;
·It is recommended to cover the balance when not in use;
·Keep the balance clean and dry.

8-1 Cleaning

·Unplug the AC adapter before cleaning;
·Do not use any aggressive cleaning agents such as solvents or
alcohols;
·Use a damp, soft cloth with mild detergent such as soap;
·Do not allow any liquids to enter the balance;
·After cleaning, wipe dry the balance with a piece of soft and dry cloth.
exceeds the limit
unstable (evaporation or
absorption of moisture)
·Close the door properly
·Remove any objects or
materials that may
obstruct the pan
·Calibrate the balance
· Press TARE key to zero
the display
·Level the balance by
turning the adjusting
feet