Ac Voltage Calibration - Datron 1062 Service Handbook

Table of Contents

Advertisement

6
1.4
AC
VOLTAGE
CALIBRATION - OPTION
10
ONLY
1.4.1
General
The
procedure
in
the table
opposite
is
all
that
is
necessary
to
completely
'AUTOCAL'
the
AC
voltage
function.
On
each
range
just a 'Zero',
'Gain'
and 'AcHf'
calibration
is
required.
If
the
'DVM
Reading
After
Calibration'
is
not
in
accordance with
the
table,
repeat operation of the
same
'CALIBRATE'
key
is
permissible to
improve
the
readings.
This
will
be necessary with the
AcHf
key.
1.4.2
Equipment
Required
A
copper
shorting
link
and
an
AC
calibration
source
e.g.
Datron
4200
Autocal
AC
Standard.
1.4.3
Checking Accuracy
after
'AUTOCAL'
To
check
the
accuracy
after
'AUTOCAL'
the
'Speci-
fication Verification' section
of the
Operating
Handbook
will
be
useful.
It
provides
tables for
quick
reference of
accuracy on
all
ranges
and
functions
in
displayed
digits.