Device Description; Purpose; Rhythmpad Data; Monitoring The Data - Cardiocity RhythmPadGP User Manual

Table of Contents

Advertisement

Ensure regular inspections of the USB cable and RhythmPad
checking for insulation and or cable damage.
No modifications of this equipment is allowed

1.11 Device Description

The RhythmPad
platform. The product contains active electrode sensors which provides a new way for
acquiring leadI ECG data from a subject's hands. The device is designed to be used
with third party software which can receive the USB formatted data from the
GP
RhythmPad
and provide analysis of the subjects cardiac rhythm, and dependent
upon the software used in conjunction with the RhythmPad
arrhythmias that may be identified and established through a leadI reading.
The full diagnostics ability of the RhythmPad
third part analysis software is restricted to the diagnostics ability of the third party
algorithm. No diagnostics capabilities are provided by the RhythmPad
GP
RhythmPad
can be expanded from a one lead to a six lead ECG acquisition tool,
when used with the 3rd electrode accessory. This six lead data can be sent to the
same third party algorithms for analysis thus providing a higher fidelity screening
system.
GP
The RhythmPad
categorized as a Class IIa device
Class 2 Electric Shock Protection
externally powered
a Type BF Applied Part
a device with an IP rating of type IPX0
not suitable for use in an Oxygen rich environment
designed for Continuous Use
a self calibrating active electrode device

1.12 Purpose

The RhythmPad
stream of ECG formatted data such that third party analysis algorithms and other
software products may receive and apply analysis to.
The
comply with instructions for installation, operation, storage and cleaning

1.13 RhythmPad Data

The RhythmPad
deciphered using LibUSB, an open source USB library.

1.14 Monitoring the Data

This User Manual details Cardiocity's RhythmPad suite of ECG analysis software.

1.15 Environmental Notice

The RhythmPad
Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided below. It
should be observed that portable and mobile RF Communications equipment can
affect MEDICAL EQUIPMENT.
RhythmPadGP Manual.doc 20/05/16
GP
is an A4 sized polycarbonate hand placement ECG monitoring
is a medical device that is:
GP
is a hand placement leadI ECG sensing system that outputs a USB
GP
RhythmPad
v1.0 is only to be used by a trained operator who
GP
emits a USB data stream in a serial data format, which can be
GP
and accessories are MEDICAL ELECTRICAL EQUIPMENT. Medical
Cardiocity Limited – RhythmPad
GP
when used in conjunction with the
GP
– User Guide- Page 10
GP
device housing
GP
, can diagnose certain
GP
alone. The

Advertisement

Table of Contents
loading

Table of Contents