Page 16
Minimum Allowance
Subsystem
Watts
Gantry
5,440
Table
200
PDU
1,000
Operator
2,165
Console
Optional Laser
800
Camera
Recommended cooling values should not be
used for calculating system input power
requirements.
** Recommended Allowance:
Growth Allowance:
* Option
LightSpeed
CT Scanner System
System Components Dimensions
BTU/hr
18,600**
700
Gantry:
3,400
7,400
2,730
Table:
7,150/24,400
9,200/31,400
Power Distribution Unit:
Operator Console:
16
With Xtream
mm
inches
Width:
2,230
87.6
Depth:
1,007
39.6
Height:
1,887
74.1
Width:
610
Length:
5,050
199
Height:
1,120
Width:
762
Depth:
585
Height:
1,270
Width:
1,238
48.7
Depth:
1,288
50.7
Height:
1,257
49.5
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Hong Kong - Fax: 852 2559 3588
GE Medical Systems-Europe

Warranty

The published Company warranty in effect on
the date of shipment shall apply. The
Company reserves the right to make changes.
All specifications are subject to change.

Regulatory Compliance

This product is designed to comply with
applicable standards under the Radiation
Control for Health and Safety Act of 1968.
24
Laser alignment devices contained within this
product are appropriately labeled according to
the requirements of the Center for Devices
44
and Radiological Health.
This product is a CE-compliant device which
satisfies regulations regarding Electro-
30
Magnetic Compatibility (EMC) and Electro-
Magnetic Interference (EMI), pursuant to IEC-
23
601.
50

Hide quick links: