Compatible Options; Siting Requirements - GE LightSpeed 16 Product Data Sheet

Ct scanner system
Table of Contents

Advertisement

g
Product Data Sheet – rev 4.2
July 4th 2003
Page 15
• 72 Hz Vertical deflection frequency
• Sync on green
• LCD Color Monitor*: 18" diagonal width
• Scan control keyboard (English language)
assembly with intercom speaker,
microphone and volume controls.
• Global modem to allow InSite connectivity
• 3-Button Mouse
• 3-Button Trackball
Image Networking:
• Standard auto-configuring
100BaseT/10BaseT Ethernet (UTP
connection)
• Direct network connection; multi-suite
ethernet card not required for gateway out of
suite
• Protocols supported:
• DICOM 3.0 network send (one IP
address at a time) and receive,
pull/query, and storage commitment
push;
• AdvantageNet (GenesisNet) point-to-
point send, receive and pull/query (no
broadcast);
• InSite point-to-point;
• TCP/IP (for system administration)

Compatible Options

• A DICOM Print Interface is standard on the
system.
• English Keyboard (B7800KE)
• French UIF & Keyboard (Z62401F)
• German UIF & Keyboard (Z62401G)
• Italian UIF & Keyboard (Z62401H)
• Spanish UIF & Keyboard (Z62401S)
• Portuguese UIF & Keyboard (Z62401P)
• ConnectPro HIS/RIS Interface with
Performed Procedure Step (PPS) M4
compatibility (B7500LN)
• Bar Code Reader (B7540RB)
• Volume Viewer (B7540LD)
• DentaScan (B7540LS)
• SmartScorePro (B7850PL) – Gating SW
• Prospective or retrospective EKG-gated
cardiac imaging in a single breath-hold
• 250-300 msec temporal resolution
* Option
LightSpeed
With Xtream
CT Scanner System
• Easy-to-use calcium scoring software
with integrated patient report generation
• Snapshot Option (B7710LS)
• CT Perfusion 2(B7700PL)
• Smart Step (B7868WL) – SW
• Smart Step HW (B7868SM & B7868SN)
• Advanced Vessel Analysis (B7868LA)
• CT Colonoscopy (B7868LC)
• CardIQ2 (B7710LR)
• Remote 21 inch Color Monitor (B7530RC)
• Exide UPS (B7999PS)
• Metal Free Table Interface for 3
Attachments (B7999PM)
• LCD Monitor (B7858LC)
• GE Color Printer (E7014LA)
• Table Tray

Siting Requirements

Minimum Suite Size:
Overall Suite:
15.5 feet (4.74 m) wide x 19.5 feet (5.96 m)
deep x 9 feet (2.75 m) high.
Recommended Suite Size:
Overall Suite:
19.5 feet (5.96 m) wide x 24.5 feet (7.49 m)
deep x 9 feet (2.75 m) high.
Exam Room: 13 feet (3.97 m) wide x 19.5
feet (5.96 m) deep.
Control Room: 11 feet (3.36 m) wide x 12.5
feet (3.82 m) deep.
Equipment Room: Not required. Power
Distribution Unit heat and noise output under
all operating conditions with all available
options is low enough to allow inclusion in the
patient scanning suite with the table and
gantry.
Temperature and Humidity:
Exam and Control Rooms: 60°-75°F (15°-
26°C) at 30%-60% relative humidity (non-
condensing).
Equipment Room: If a separate equipment
room is used to house the PDU, the allowable
temperature range is 60°-84°F (15°-29°C) at
30%-60% relative humidity (non-condensing).
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:
16
Tokyo, Japan - Fax: 81 425 85 5490
Hong Kong - Fax: 852 2559 3588
GE Medical Systems-Europe
Temperature Rate of Change: 3°C/hour max.
Relative Humidity Rate of Change: 5%
RH/hour max.
Power Requirements:
The only facility input to the system is a 380 to
480 V nominal, 3 phase Delta or Wye, 50/60
Hz, 90 kVA service, 20 kVA average power;
main disconnect to be located within 5 feet
(1.5 m) of the PDU. The facility must also
provide a protective disconnect device with
low voltage, low energy local and multi-point
remote capability, in the line feeders to the
PDU.
rd
Party
Complete, detailed specifications of all power
requirements are available upon request. For
most installations, the Lightspeed
Scanner System does not require any power
conditioning equipment to be used in
conjunction with the PDU. Regulators are not
recommended for use with this system. For
those sites with known large power line
transients, a suppresser filter for the system
computer and peripherals may be useful. In
general, suppresser filters are not
recommended.
Compatible UPS options are available if
desired.
Cooling Requirements:
The cooling requirements do not include
cooling for the room lighting, personnel or
non-CT equipment present. Cooling
requirements are listed by subsystem to allow
planning for each room of the CT suite.
Cooling requirements are given for minimum,
recommended and growth allowance
scenarios.
• The minimum cooling figures assume
patient throughput of 3 patients per hour and
75 scan rotations per patient.
• The recommended cooling requirements
assume patient throughput limited by the
tube cooling algorithm.
• The suite cooling can be sized for future
developments by using the growth
allowance figures. This cooling will
accommodate more patients per hour and/or
potential future system enhancements.