Insert the thermocouples and connect them to the terminal blocks. The leg of the thermocouple
with a blue paint mark is the negative. The negative side of the connecting cable is marked with a
double number (e.g. 11).
2.6
Electrical Connections
Connection by a qualified electrician is recommended.
The /300 models are designed only for single phase electrical supplies with or without neutral.
The larger models may be made for 1- or for 3-phase use. The furnace must be connected only to
the type and voltage of supply for which it was ordered.
The voltage or range of voltages on which the furnace may be operated is given on the furnace
rating label. Check that the supply voltage is compatible with the voltage on the label, and that the
current capacity is sufficient for the amperage on the label, before connection to the supply. A
table of the most common ratings is given in section 8.1 of this manual.
The supply point must be within reach of the furnace operator and must incorporate either an
isolating switch which operates on all live conductors, or a quickly removable plug.
The supply MUST incorporate an earth (ground).
CONNECTION DETAILS
Supply
1-phase
Supply
3-phase
2.7
Inserting a Tube
If a work tube is supplied with the furnace, then the hole through the insulation is correctly sized.
Carefully line up the tube and ease it through the chamber. Using a pole such as a broom handle
may help with this operation.
For a tube of customer supply, ensure that there is approximately 1mm of free play sideways and
upwards. If the tube is too tight it will crack when heated. A tight hole can be eased by rotating the
tube, which has a sufficiently abrasive surface. Remove loose dust with a domestic vacuum
cleaner. Safety note: see section 5.2.
If the holes in the insulation are too large, please
consult Carbolite for technical assistance.
If a long tube is specified, then a pair of
cylindrical end-guards is supplied. Bolt these
onto the furnace ends. The guards incorporate a
hookbolt and strap arrangement for supporting
the tube, to prevent failure from mechanical
stress. These should be adjusted to take the
weight of the tube without lifting it clear of the
insulation.
2.8
4
Terminal label
Cable colour
L
Brown
N
Blue
PE
Green/Yellow
Terminal label
Cable colour
L1
Black
L2
Black
L3
Black
N
Light Blue
PE
Green/Yellow
supply type
Live-Neutral
Reversible or Live-Live
To live
to either power conductor
To neutral
to the other power conductor
To earth (ground)
to earth (ground)
to phase 1
to phase 2
to phase 3
to neutral except delta
to earth (ground)
MF32