10.2VDC. More than 5,000 scale divisions have been selected in Err d Ntep or Angle mode. More than 20,000 scale divisions have been selected in NO mode. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
Loop Back 0 Fastest to 7 FILTER slowest (3) SPAN POINT Cal Factors ? 1 TO 99 YES, NO MOTION BAND Divisions (2) SPAN POINT DEFAULT 0 to 99 MOTION DELAY UPDATES (4) CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
ZERO-------KEY IS USED TO NAVIGATE UP. GRS/NET---KEY IS USED TO NAVIGATE DOWN. TARE--------KEY IS USED TO NAVIGATE LEFT. 0 - 9 Deg. ANGLE LIMIT lb/kg---------KEY IS USED TO NAVIGATE RIGHT. PRINT-------KEY IS USED TO ENTER DATA AND RETURNS. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
Dashes: When the scale is in motion the display will show all Dashes. S10 Zero Band 1 to 99 divisions (2) the weight display must return to Zero within the selected number of divisions to to be considered Zero. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
(0) to 30 minutes. The display will turn off after the the set amount of time elapses with no scale activity. Angle Limit (6) Adjusts maximum angle. Exits back to the main Menu. ( ) indicates Factory Set defaults. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
Press the key when finished to enter lb/kg PRINT the data. The display will return to . Exit Set up Mode by pressing the Key, End Tare will be displayed, then press Print/Enter. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
If the displayed weight does not match the known test weight, follow the steps described for C3 Span point on adjusting the weight and entering the value. Exits back to the main Menu. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
After calibration is complete you may also return to these correction points and make changes to its value without affecting any of the original calibration points. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
2 - Slave Display Output (alphanumeric) continuous Stx GR or NT or TR, six ASCII Characters (indicated weight), lb or kg, CR, LF [15 bytes total output]. 3 - RF Link Output. 4 - Used for QSI terminal. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
Page 13
Receives R3.3 RF link output string displaying data as it appears on the scale and allows full control of all scale meter functions. Exits back to the main Menu. ( ) indicates Factory Default setting CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
"r you sure?" will be displayed then press the Tare key to exit without Defaulting, or press the Print key to reset the meter to factory default. Exits back to the main Menu. CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...
CAMBRIDGE, the equipment has not been mechanically, environmentally, or electrically abused. This warranty is limited, at the option of CAMBRIDGE, to repair, replace or an appropriate credit adjustment not to exceed the original equipment sale price paid to CAMBRIDGE.
7.0 Assistance If at any time you require assistance with your Model: CSW-10AT Meter please contact us at: CAMBRIDGE SCALE WORKS, INC. 115 West Mary Street Cumberland, MD 21502 Phone: (301) 724-4082 Fax: (301) 724-4964 CAMBRIDGE SCALE WORKS, INC. MANUAL P/N 5999-1014-01...