O
PERATION
12
Sensor Break Protection
This controller provides sensor break protec-
tion in the event the thermocouple opens. If an
open thermocouple condition occurs, the digital
display will indicate 5 degrees or less and the
power to the heating element will be shut off
(CYCLE light will extinguish).