GUIDANCE AND MANUFACTURER'S DECLARATION
Immunity test
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radio, AM and FM radio broadcast, and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the ReliOn
6021RELN2 is used exceeds the applicable RF compliance level above, the ReliOn
6021RELN2 should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as reorienting or relocating
the ReliOn
6021RELN2.
TM
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IEC 60601
Compliance
Test level
level
Electromagnetic environment
– guidance
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol: