0 to +-200v amplitude, 350 khz bandwidth high-voltage function generator and variable-gain linear amplifier with ieee 488.2 and rs-232 control (39 pages)
0 to ±400v amplitude, 50 khz bandwidth high-voltage function generator and variable-gain linear amplifier with ieee 488.2 and rs-232 control (40 pages)
Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item. This warranty does not apply to units which have been dissembled, modified or subjected to conditions exceeding the applicable specifications or ratings.
TABLE OF CONTENTS WARRANTY........................2 TECHNICAL SUPPORT....................2 TABLE OF CONTENTS....................3 INTRODUCTION.......................5 HIGH-VOLTAGE PRECAUTIONS..................5 SPECIFICATIONS......................6 REGULATORY NOTES....................7 FCC PART 18..........................7 EC DECLARATION OF CONFORMITY..................7 DIRECTIVE 2002/95/EC (RoHS)....................8 DIRECTIVE 2002/96/EC (WEEE)....................8 INSTALLATION........................9 VISUAL CHECK........................9 POWER RATINGS........................9 CONNECTION TO THE POWER SUPPLY................9 PROTECTION FROM ELECTRIC SHOCK................10 ENVIRONMENTAL CONDITIONS..................11 FUSES..........................12 AC FUSE REPLACEMENT....................12...
The input range is 0 to ±2V, and the output range is 0 to ±200V. The outputs can drive load impedances of 50 kΩ and higher. The amplifier bandwidth is DC-100 kHz for model AV-110B-PS-D, and DC-350 kHz for model AV-110G-PS-D.
If interference is observed, check that appropriate well-shielded cabling is used on the output connectors. Contact Avtech (info@avtechpulse.com) for advice if you are unsure of the most appropriate cabling. Also, check that your load is adequately shielded. It may be necessary to enclose the load in a metal enclosure.
Directive 2002/95/EC. DIRECTIVE 2002/96/EC (WEEE) European customers who have purchased this equipment directly from Avtech will have completed a “WEEE Responsibility Agreement” form, accepting responsibility for WEEE compliance (as mandated in Directive 2002/96/EC of the European Union and local laws) on behalf of the customer, as provided for under Article 9 of Directive 2002/96/EC.
INSTALLATION VISUAL CHECK After unpacking the instrument, examine to ensure that it has not been damaged in shipment. Visually inspect all connectors, knobs, and handles. Confirm that a power cord is with the instrument. If the instrument has been damaged, file a claim immediately with the company that transported the instrument.
5. Do not attempt any repairs on the instrument, beyond the fuse replacement procedures described in this manual. Contact Avtech technical support (see page 2 for contact information) if the instrument requires servicing. Service is to...
ENVIRONMENTAL CONDITIONS This instrument is intended for use under the following conditions: 1. indoor use; 2. altitude up to 2 000 m; 3. temperature 5 °C to 40 °C; 4. maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % relative humidity at 40 °C;...
FUSES This instrument contains four fuses. All are accessible from the rear-panel. Two protect the AC prime power input, and two protect the internal DC power supplies. The locations of the fuses on the rear panel are shown in the figure below: Fuses #1 and #2 Fuse #4 Fuse #3...
DC FUSE REPLACEMENT The DC fuses may be replaced by inserting the tip of a flat-head screwdriver into the fuse holder slot, and rotating the slot counter-clockwise. The fuse and its carrier will then pop out. FUSE RATINGS The following table lists the required fuses: Recommended Replacement Part Nominal Fuses...
FRONT PANEL CONTROLS 1. POWER Switch . This is the main power switch. When turning the instrument on, there may be a delay of several seconds before the instrument appears to respond. 2. OVERLOAD Indicator . When the instrument is powered, this indicator is normally green, indicating normal operation.
Page 15
this dial can be used to enable/disable this feature. 6. OUT A Connector . This BNC connector provides the main output of the Channel A amplifier. The output is an amplified version of the input on (3). The gain (V ) is controlled by (4).
REAR PANEL CONTROLS 1. AC POWER INPUT . An IEC-320 C14 three-pronged recessed male socket is provided on the back panel for AC power connection to the instrument. One end of the detachable power cord that is supplied with the instrument plugs into this socket. 2.
GENERAL INFORMATION BASIC CONTROL The AV-110B-PS-D is a DC-100 kHz variable-gain linear amplifier, and the AV-110G- PS-D is the same but with an extended bandwidth range of DC-350 kHz. The gain of each of the two channels (designated “A” and “B”) is variable from +1 to +100, and is adjusted by rotating the “Gain”...
OPERATIONAL CHECK This section describes a sequence to confirm the basic operation of the instrument. It should be performed after receiving the instrument. It is a useful learning exercise as well. SYNC OUTPUT SIGNAL SOURCE (±2V maximum) MAIN OUTPUT AV-110G-PS-D REAL-TIME AMPLIFIER OSCILLOSCOPE...
Page 19
5) Set the gain control to minimum (0.0). Turn on the amplifier and the signal generator. 6) Rotate the gain control to its maximum setting. The Channel 2 waveform should increase to ±200V, and have a shape similar to that of the Channel 1 waveform. 7) Repeat the previous steps using the IN B and OUT B connectors.
There are no user-adjustable internal circuits. For repairs other than fuse replacement, please contact Avtech (info@avtechpulse.com) to arrange for the instrument to be returned to the factory for repair. Service is to be performed solely by qualified service personnel.
MAINTENANCE REGULAR MAINTENANCE This instrument does not require any regular maintenance. On occasion, one or more of the four rear-panel fuses may require replacement. All fuses can be accessed from the rear panel. See the “FUSES” section for details. CLEANING If desired, the interior of the instrument may be cleaned using compressed air to dislodge any accumulated dust.