Electrical Specifications - SEFRAM 7205 User Manual

6000 counts trms ac digital multimeters
Table of Contents

Advertisement

Available languages

Available languages

Pollution degree : 2
Safety : Complies with EN61010-1,CAT.IV. 600V, CAT.Ⅲ. 1000V
EMC : EN 61326-1
Shock vibration : Sinusoidal vibration per MIL-PRF- 28800F
(5 ~ 55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on concrete floor.
Indoor Use.

Electrical Specifications

Accuracy is given as ±(% of reading + counts of least significant
digit) at 23°C ± 5°C, with relative humidity Less than 80% R.H.
1. Temperature coefficient
0.1 x (Specified accuracy) / °C, < 18°C, > 28°C
2. AC Function
ACV and ACA specifications are ac coupled, true R.M.S. The crest
factor may be up to 3.0 as 4000 counts.
For square wave, Accuracy is unspecified.
For non-sinusoidal waveforms, Additional Accuracy by Crest
Factor (C.F.) : Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
3. DC mV
Range
OL Reading
Resolution
600.0mV
660.0mV
0.1mV
Input Impedance : 10MΩ
Overload Protection : AC/DC1000V
21
Accuracy
±(0.5% + 5D)
M 7205-6-7 M 00

Advertisement

Table of Contents
loading

This manual is also suitable for:

72067207