GLI International C53 Operating Instructions Manual

Conductivity analyzer
Table of Contents

Advertisement

Back to MANUALS Menu
OPERATING INSTRUCTION MANUAL
Worldwide Headquarters and Sales:
GLI International, Inc.
9020 West Dean Road
Milwaukee, Wisconsin 53224
U.S.A.
In the interest of improving and updating its equipment, GLI reserves the right to alter specifications to equipment at any time.
Rev. 5-1000
GLI_CondAnalyzer_C53_om_D500
Model C53
Conductivity Analyzer
(Universal-mount 1/2 DIN style;
selectable for conductivity, resistivity, TDS,
and calculated Sensor A and B measurement)
Phone:
[414] 355-3601
Fax:
[414] 355-8346
E-mail:
info@gliint.com
Web:
www.gliint.com
A
Viridor Instrumentation
Represented By:
company
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the C53 and is the answer not in the manual?

Questions and answers

Summary of Contents for GLI International C53

  • Page 1 Milwaukee, Wisconsin 53224 Web: www.gliint.com U.S.A. In the interest of improving and updating its equipment, GLI reserves the right to alter specifications to equipment at any time. Viridor Instrumentation company Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 2 Back to MANUALS Menu Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 3: Important Safety Information

    FOR PERSONAL INJURY. CAUTION: A CAUTION LOOKS LIKE THIS. IT ALERTS YOU TO POSSIBLE INSTRUMENT MALFUNCTION OR DAMAGE. NOTE: A note looks like this. It alerts you to important, useful operating information. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 4 WARRANTY GLI International, Inc. warrants the Model C53 to be free from defects in material or workmanship for a period of 2 years (24 months) from the date of shipment of this product from our facility. A warranty claim will not be honored if defects are...
  • Page 5 “wet” calibrated. However, since measured conductivity is greatly affected by small changes in temperature, GLI strongly recommends using its DRY-CAL method for highest measuring accuracy of conductivity and temperature. Besides, DRY-CAL is actually a normal part of Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 6 D. With the “SENSOR A” line selected, press ENTER key to display Ø Ø E. Use key to select the “ZERO” line, and press ENTER key to display (continued on next page) Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 7 GLI-certified “CELL K” value. (Use keys to × × Ø Ø coarse adjust, and keys to fine adjust.) Then press ENTER key to enter the value. (continued on next page.) Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 8 To further configure the analyzer to your application requirements, use the appropriate CONFIGURE screens to make selections and “key in” values. Refer to PART THREE, Section 4 for complete con- figuration details. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 9: Table Of Contents

    2.1 Displaying Main Branch Selection Screen ..........35 2.2 Displaying Top-level Menu Screens..........35-36 2.3 Displaying Submenu Screens ............... 36 2.4 Adjusting Edit/Selection Screen Values ..........36 2.5 Entering (Storing) Edit/Selection Screen Values/Choices..... 36 Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 10 5.1 Things to Know About Calibration ............62-63 5.2 Zero Procedure (first-time sensor calibration only) ......63-64 5.3 DRY-CAL Method (highly recommended): SELECT CELL K (sensor’s GLI-certified “K” value) ....64-65 SET T FACTOR (sensor’s GLI-certified “T” factor) .....65-67 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 11 Assigning User-defined Date ............89-90 Viewing Identification (ID) ..............90 Viewing Revisions................90 8.7 “Master Reset” Function................ 91 8.8 “Refresh” Function ................91 8.9 Protocol Command Set for PC Programming ........91 Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 12 P A R T F I V E - S P A R E P A R T S A N D A C C E S S O R I E S ........................99 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 13 Analyzer Configuration Settings (Ranges/Choices and Defaults) ........59-61 Table D Conductivity Reference Solutions ..................68 Table E Relay Overfeed Timer Interactions with Other Analyzer Functions........81 Table F Full-scale Equivalent Test Resistance Values..............96 Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 14 Back to MANUALS Menu Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 15: Part O N E - I N T R O D U C T I O N

    However, the analyzer can be traditionally “wet” calibrated. For calibration details, re- fer to PART THREE, Sections 5.2 and 5.3 or 5.4. Each analog output mA value can also be calibrated (Section 5.5). Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 16 CALCULATION. NOTE: When a relay is set to function as a STATUS relay, it is no longer configurable. Instead, it becomes a Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 17: Modular Construction

    1.4 Analyzer A label with the analyzer model number, serial number, Serial Number build date, and other items is affixed to the top of the enclo- sure. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 18: Emi/Rfi Immunity

    US standards and meets European IEC 801-series testing for electromagnetic and radio fre- quency emissions and susceptibility. Refer to Figure 1-1 and the specifications in Section 2.1 for more information. FIGURE 1-1 EMI/RFI Immunity Diagram Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 19: Specifications

    NOTE: Depending on Sensor A or B’s selected measurement, only some of the following temperature compensation methods are available: • Linear % per °C slope • Built-in ammonia temperature properties table • Built-in natural water temperature properties table • No compensation Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 20: Section 2 Specifications

    1/2 inch (13 mm) conduit holes; nylon mount- ing bracket, and stainless steel hardware Mounting Configurations..... Panel, surface, and pipe (horizontal and vertical) mounting Net Weight ......... 3.5 lbs. (1.6 kg) approximately Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 21: Part T W O - I N S T A L L A T I O N

    Figure 2-1 illustrates the various ways to mount the ana- lyzer using the supplied bracket and hardware. Determine the mounting method and attach the hardware as shown in the respective illustration. Refer to Figure 2-2 for analyzer installation dimension details. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 22: Figure 2-1 Analyzer Mounting Arrangements

    Back to MANUALS Menu PART TWO - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS FIGURE 2-1 Analyzer Mounting Arrangements Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 23: Conduit Hole Requirements

    3H1230, for cable entries.) Seal unused cable entry holes with appropriate plugs. NOTE: Use NEMA 4-rated fittings and plugs to maintain the watertight integrity of the NEMA 4X enclosure. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 24: Electrical Connections

    For installations where the distance between sensor and analyzer exceeds the sensor cable length, indirectly connect the sensor to the analyzer using a junction box and interconnect cable. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 25: Figure 2-3 Terminal Designations For Analyzers With "B" Prefix Serial Number

    180-260 VAC 10VA 50/60 Hz 80mA 100mA FIGURE 2-3 Terminal Designations for Analyzers with “B” Prefix Serial Number FIGURE 2-4 Terminal Designations for Analyzers with “A” or No Letter Prefix Serial Number Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 26: Figure 2-5 Connecting Gli Contacting Conductivity Sensor(S) To Analyzers With "B" Prefix Serial Number

    FIGURE 2-5 Connecting GLI Contacting Conductivity Sensor(s) to Analyzers with “B” Prefix Serial Number FIGURE 2-6 Connecting GLI Contacting Conductivity Sensor(s) to Analyzers with “A” or No Letter Prefix Serial Number Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 27: Analog Outputs

    Output 1 remains available for normal use. However, in a HART multi-drop wiring configu- ration, Output 1 becomes dedicated to that function and cannot be used. See PART THREE, Section 8 for more HART communication information. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 28: Relay Outputs

    WARNING: MAKE SURE LINE POWER IS NOT PRESENT WHILE CONNECTING WIRES TO TB2 RELAY TERMINALS. FIGURE 2-7 Connecting Control/Alarm Device(s) To Electromechanical Relay(s) Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 29: Closed Contact Input

    NOTE: For 230 volt split phase line power, be sure to con- form to local codes with regard to fusing the 115 volt line connected to the “N” terminal. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 30: Figure 2-9 Connecting 115 V Single Phase To Analyzers With "A" Or No Letter Prefix Serial Number

    FIGURE 2-12 FIGURE 2-13 Connecting 230 Volt Split Phase Connecting 230 Volt Split Phase to Analyzers with “B” Prefix Serial Number to Analyzers with “A” or No Letter Prefix Serial Number Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 31: Part T H R E E - O P E R A T I O N

    When a relay overfeed timer is used and it has “timed out,” the respective relay annunciator con- tinuously blinks until the overfeed condition is resolved. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 32: Keypad

    ESC key once takes the display up one level to the MEASURE screen.) This key can also “abort” the pro- cedure to change a value or selection. FIGURE 3-1 Analyzer Keypad Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 33: Measure Screen (Normal Display Mode)

    The MEASURE screen can show four different readout ver- sions. To select between them, in continuous loop Ö Ö Õ Õ key . These are examples of sequence, press the the different versions: Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 34 When the measured value is beyond the analyzer measur- ing range, a series of “ + ” or “ - ” screen symbols appear, respectively indicating that the value is above or below range. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 35: Menu Structure

    Menu Structure Tip! The symbol pointing at each listed item indicates there is a related lower-level submenu screen, sub-submenu screen or edit/selection screen. Some menu lists are too long to completely fit on the Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 36: Displaying Submenu Screens

    (store) it into the non-volatile analyzer memory. Values/Choices The previous screen will then re-appear. NOTE: You can always press the ESC key to abort saving a new setting. The original setting will be retained. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 37: Display Contrast Visibility

    With the desired language displayed, press ENTER key to enter this selection. NOTE: After a language is selected and entered, all screens will be displayed in that language. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 38: Configuring Sensor (A And B) Characteristics

    • CONDUCTIVITY: Selects conductivity measurement. • RESISTIVITY: Selects resistivity measurement. • TDS: Selects total dissolved solids measurement. WARNING: WHEN THE MEASUREMENT IS CHANGED, ALL USER-ENTERED VALUES ARE REPLACED BY FACTORY-DEFAULT VALUES. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 39: Select Display Format

    XX.XX M Ω • cm 2.000 µS/cm 2.000 mS/cm XXX.X K Ω • cm 20.00 µS/cm 20.00 mS/cm 200.0 µS/cm 200.0 mS/cm 2000 µS/cm 2000 mS/cm For TDS XXXX ppm XXXX ppb Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 40: Select Temperature Compensation

    “CONFIG LINEAR Temperature Compensation” for details. 3. With the desired choice displayed, press ENTER key to enter this selection. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 41: Config Tds Measurement

    ENTER key to enter the value. 6(7 )$&725 (;,7 C. After the screen re-appears, press ESC key once to return to the “SENSOR A” submenu screen. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 42: Config Linear Temperature Compensation

    “SET REF TEMP” line. 5. Press ENTER display screen like . Use arrow keys to adjust the dis- played value to the desired reference temperature, and press ENTER key to enter it. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 43: Select Cell K (Sensor's Gli-Certified "K" Value)

    GLI-certified “K” value, and press ENTER key to enter the value. 6. After the screen re-appears, press ESC key once to return to “SENSOR A” submenu screen. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 44: Set Filter Time

    2. Press display screen like Ø Ø × × keys to view both . Use choices (OFF or ON). With the desired choice dis- played, press ENTER key to enter this selection. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 45: Enter Note (Top Line Of Measure Screen)

    NOTE: When a temperature element type has been se- lected but the element is not connected to the analyzer, a “WARNING CHECK STATUS” message will appear. To prevent or clear this message, con- nect the element or select “MANUAL.” Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 46 B. Press ENTER key to display a screen like . Use arrow keys to adjust the dis- played value to the desired fixed temperature, and press ENTER key to enter it. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 47: Set T Factor (Sensor's Gli-Certified "T" Factor)

    3 ° C less than the known solution temperature you would decrease the sensor T FACTOR by subtracting 11.55 from it. *Sensors with a PT 100 temperature element provide inherently less accurate temperature readings and are not recommended. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 48: Configuring Calculated Sensor A And B Measurement

    (NONE, % REJECT, % PASS, RATIO A/B, RATIO B/A, DIFF A-B or DIFF B-A). With the desired choice displayed, press ENTER key to en- ter this selection. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 49: Select Display Format

    2. Press ENTER display screen like Ø Ø × × . Use keys to view both choices (°C or °F). With the desired choice displayed, press ENTER key to enter this selection. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 50: Configuring Analog Outputs (1 And 2)

    % passage, ratio A/B, ratio B/A, diff A-B or diff B-A). *An analog output can only represent the calculated measurement when two sensors are used and the analyzer has been correctly configured for CALCULATION. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 51: Set 0/4 Ma And 20 Ma Values (Range Expand)

    20 mA is desired, and press ENTER key to enter the value. NOTE: If the same values are set for 0/4 mA and 20 mA, the output automatically goes to, and remains at, 20 mA. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 52: Set Transfer Value (Ma)

    “SET FILTER” line. ENTER 2. Press display screen like . Use arrow keys to adjust the dis- played value to the desired filter time, and press ENTER key to enter it. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 53: Select Scale 0 Ma/4 Ma (Low Endpoint)

    XFER or ACTIVE during calibration, they remain in that cor- responding on/off state regardless of the connection status at the TTL input terminals. These instructions configure Relay A. Configure other relays in the same way using their respective menu screens. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 54: Set Parameter (Representation)

    Each relay can be selected to function as a: SET FUNCTION Mode (alarm, control or status) • ALARM relay (with separate high and low alarm points and deadbands) that operates in response to the selected measured value. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 55: Set Transfer Mode (Relay On Or Off)

    2. Press ENTER key to display . Use Ø Ø × × keys to view both choices (DE-ENERGIZED or ENERGIZED). With the desired choice displayed, press ENTER key to enter this selection. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 56: Activation (Configuration Values)

    “low” values are lower than “high” values. When a relay is set for STATUS function, the symbol at the start of the “ACTIVATION” line de- notes that this menu item is not available. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 57 4. Use the same basic keypad operations described in previous setup procedures to enter the desired value for the displayed relay activation setting. 5. Repeat this procedure for each relay activation setting. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 58: Set Passcode (Feature Enabling Or Disabled)

    ENTER display screen like Ø Ø × × . Use keys to view both choices (DISABLED or ENABLED). With the desired choice displayed, press ENTER key to enter this selec- tion. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 59: Configuration Setting Summary (Ranges/Choices And Defaults)

    DIFF: Same defaults as those listed “Sensor A and B Configuration in the “Sensor A and Sensor B Con- Settings” category above figuration Settings” category above (Table C continued on next page.) Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 60 TDS: 0 ppm or 0 ppb TEMP: -20.0 to +200.0 ° C or -4.0 to 392.0 ° F TEMP: 0.0 ° C or 32.0 ° F (Table C continued on next page.) Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 61: Table C Analyzer Configuration Settings (Ranges/Choices And Defaults)

    A, TEMPERATURE B or CALCULATED SIM SENSOR? Same ranges as those listed for Present measured value of “SET 4 mA VALUE” function in the Sensor A’s selected parameter. “OUTPUT Configuration Settings” category. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 62: Analyzer Calibration

    ESC key. After “ABORT: YES?” screen appears, do one of the following: • Press ENTER key to abort. After the “CONFIRM ACTIVE?” screen appears, press ENTER key Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 63: Zero Procedure (First-Time Sensor Calibration Only)

    4. With the “SENSOR A” line selected, press ENTER key to display Ø Ø 5. Use key to select the “ZERO” line, and press ENTER × × Ø Ø key to display . Use key to view Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 64: Dry-Cal Method (Highly Recommended)

    “SELECT CELL K” and “SET T FACTOR.” &21),*85( 7(670$,17 (;,7 1. Press MENU key to display SELECT CELL K (sensor’s GLI-certified “K” value) Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 65: Set T Factor (Sensor's Gli-Certified "T" Factor)

    GLI-certified “K” value, and press ENTER key to enter the value. SET T FACTOR After entering the “K” value, complete DRY-CAL calibration (sensor’s GLI-certified by entering the GLI-certified “T” factor: “T” factor) Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 66 3 ° C less than the known solution tem- perature you would decrease the sensor T FACTOR by subtracting 11.55 from it. *Sensors with a PT 100 temperature element provide inherently less accurate temperature readings and are not recommended. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 67: Point Sample Method (Wet Calibration)

    NaCl to one liter of high purity, de-ionized, CO -free water that is 25°C to obtain the listed conductivity. Solution con- ductivity can be decreased by dilution with de- ionized water. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 68: Table D Conductivity Reference Solutions

    2. Press MENU key to display 3. With the “CALIBRATE” line selected, press ENTER key 6(1625 % 2873876 (;,7 to display 4. With the “SENSOR A” line selected, press ENTER key to display Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 69 ENTER key again to return the analog outputs and re- lays to their active states (MEASURE screen appears). This completes “1 POINT SAMPLE” calibration. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 70: Analog Outputs (1 And 2) Calibration

    The displayed value is “counts” -- not mA -- that dy- namically change when the output is adjusted. 6. Use a calibrated digital multimeter to measure Output 1’s actual minimum value provided at Terminals 2 and 3 on TB1. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 71 “20.00 mA” on the digital multimeter -- not the analyzer display, and press ENTER key to com- plete calibration of the maximum endpoint value. This completes analog Output 1 calibration. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 72: Test/Maintenance

    Ø Ø key to select the “TEST/MAINT” line. 2. Press ENTER key to display 3. With the “STATUS” line selected, press ENTER key to display the “STATUS: ANALYZER OK” screen. This Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 73 (Relay energized; LED is on.) NOTE: COUNTING only applies to control relays. 6. To end status checking, press ESC key or ENTER key (display returns to TEST/MAINT top-level menu screen). Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 74: Hold Outputs

    Ø Ø 1. With the screen displayed, use to select the “OVERFEED RESET” line. 2. Press ENTER key to display “OVERFEED RESET” screen. To reset all overfeed timers at once, press Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 75: Output (1 And 2) Analog Test Signals

    Relays A, B, C, and D can be tested to confirm their opera- 6.5 RELAY (A, B, C, and D) tion. These instructions only test Relay A. Test other Operating Test relays in the same way using their respective menu screens. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 76: Eprom Version Checking

    “EPROM VERSION” line. 2. Press ENTER key to view the EPROM version screen. 3. To return to the TEST/MAINT top-level menu screen, press ESC key or ENTER key . Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 77: Select Sim Measurement

    TEST/MAINT top-level menu screen. 6.8 SIM SENSOR Setting After selecting the type of simulated measurement (subsec- tion 6.7), set the desired simulation value. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 78: Reset Defaults (Configuration Settings To Factory-Set Values)

    2. Press ENTER key to display the “RESET DEFAULTS: ARE YOU SURE?” screen, asking if you really intend to perform this extreme action. (If you want to abort this action, press ESC key now.) Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 79 3. Press ENTER key to reset stored configuration and calibration settings factory defaults (“RESET DEFAULTS: DONE” screen appears, acknowledging reset has occurred). 4. To return to the TEST/MAINT top-level menu screen, press ESC key or ENTER key . Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 80: Relay Overfeed Timer Feature

    A relay overfeed timer can, and often will, interact with other analyzer functions while those functions are in use. Table E Analyzer Functions on the next page explains common overfeed timer interac- tions. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 81: Table E Relay Overfeed Timer Interactions With Other Analyzer Functions

    (or a value you On relay is turned Overfeed timer simulate) causes the relay to turn off. “off” by simulation value was timed out Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 82: Hart Option

    Fax: [512] 794-8893 Website: www.hartcomm.org For information on the HART Communicator Model 275, contact: Fisher-Rosemount Systems 12000 Portland Avenue South Burnsville, Minnesota 55337-1535 USA Headquarters: [612] 895-2000 Service: [800] 654-7768 Fax: [612] 895-2244 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 83: Analyzer Operating Modes For Hart Network

    HART-capable software (or software including GLI Device Specific Command sets). The GLI analyzer never initiates a command sequence, but always responds to commands from the master. Up to two master devices may be Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 84: Single Mode (Point-To-Point) Wiring Arrangement

    Figure 3-3 and connect all devices, including up to two masters, to the 4-20 mA analog Output 1 signal. FIGURE 3-3 HART SINGLE MODE (Point-to-Point) Wiring Connections (for single analyzer) Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 85: Multi-Drop Wiring Arrangement

    Output 1 signal, matching polarity as shown. 4. Up to two masters can be connected to the 4-20 mA analog Output 1 signal cable. FIGURE 3-4 HART MULTI-DROP Wiring Connections (for multiple analyzer network) Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 86: Hart Preferences Setup

    Required Preambles ber of preambles required by the analyzer from the master. 1. With the “GLI SETUP” top-level menu screen dis- played, select the “HART INFO” line and press key . Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 87: Device Preferences Setup

    The “Model” information screen shows the model number of Model Number the analyzer. 1. With the “GLI SETUP” top-level menu screen dis- played, select “DEVICE INFO” line and press key . Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 88: Viewing Manufacturer

    4. Press the F4 key to enter the tag, and the F2 key to send the tag to the analyzer. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 89: Assigning A Descriptor

    “DEVICE INFO” line and press key . 2. With the “DEVICE INFO” submenu screen displayed, key to display its select the “Date” line and press the related information screen. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 90: Viewing Identification (Id)

    3. With the “DEVICE REVISION” sub-submenu screen displayed, select the appropriate line and press the key to display its related information screen. 4. Press the F4 key to return to the “DEVICE INFO” sub- menu screen. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 91: Master Reset" Function

    Set for PC limited operability. The Device Specific Command set for all Programming existing GLI analyzers is available on request for creating a full-featured HART-capable program to run on an IBM- compatible PC. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 92: Part F O U R - S E R V I C E A N D M A I N T E N A N C E

    -- or -- • Replace the complete scaling board assembly contain- ing the relays. Refer to PART FIVE -- Spare Parts -- for the GLI scaling board assembly part number. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 93: Preserving Measurement Accuracy

    Cut cables to proper length during installation to avoid unnecessary inductive pickup (“electrical noise” may interfere with sensor signal). Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 94: Troubleshooting

    3. Note the analyzer reading now and compare it with the reading taken in step 1. If the reading changed, a ground loop exists. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 95: Finding Source Of Ground Loop

    1% tolerance, 1097 ohm resis- tor between Terminals 18 (blue) and 19 (white) on TB1. (When configured for a PT100 temperature element, use a 1% tolerance, 110 ohm resistor.) Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 96: Verifying Interconnect Cable Integrity

    Verifying Interconnect If these readings are achieved, the analyzer is operating Cable Integrity properly, but the interconnect cable (if used) may be faulty. WARNING: DISCONNECT LINE POWER AVOID POSSIBILITY OF ELECTRICAL SHOCK. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 97 If the reading is achieved, the interconnect cable and/or junction box connections are probably faulty. Use a digital multime- ter to check the interconnect cable for shorted or open wires. Rev. 5-1000 Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
  • Page 98: Analyzer Repair/Return

    (Recommendation: Use the original GLI shipping carton or an equivalent.) Also, GLI will not accept analyzers returned for re- pair or replacement unless they are thoroughly cleaned and all process material is removed. Model C53 Conductivity Analyzer (universal-mount 1/2 DIN) Rev. 5-1000...
  • Page 99 Serial Number Power Supply/Scaling Board Assembly..C53G1040-101 Ribbon Interconnect Cable......4W001 The following parts are common to all C53 Analyzers regardless of serial number prefix: Fuse Kit (one 80 mA fuse and one 100 mA fuse per package)......1000G3315-101 Mounting Hardware Kit ........1000G3228-101 Analyzer Sun Shield ........1000G3088-001...

Table of Contents