BIRD Site Analyzer SA-2000 Operating Instructions Manual

Cable and antenna tester for wireless systems
Table of Contents

Advertisement

SITE ANALYZER™
CABLE AND ANTENNA TESTER
FOR WIRELESS SYSTEMS
OPERATING INSTRUCTIONS
©Copyright 1999, 2002 by Bird Electronic Corporation
Instruction Book Part Number 920-7002A800 Revision K
Site Analyzer is a trademark of Bird Electronic Corporation
Microsoft® and Windows® are registered trademarks
of the Microsoft Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Site Analyzer SA-2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for BIRD Site Analyzer SA-2000

  • Page 1 CABLE AND ANTENNA TESTER FOR WIRELESS SYSTEMS OPERATING INSTRUCTIONS ©Copyright 1999, 2002 by Bird Electronic Corporation Instruction Book Part Number 920-7002A800 Revision K Site Analyzer is a trademark of Bird Electronic Corporation Microsoft® and Windows® are registered trademarks of the Microsoft Corporation...
  • Page 2 This page is not blank...
  • Page 3: Safety Precautions

    Safety Precautions Remove Power Observe general safety precautions. Do not open the instrument with the power on. Safety Earth Ground An uninterruptible safety earth ground must be supplied from the main power source to the instrument. Grounding one conductor of a two conductor power cable is not sufficient protection.
  • Page 4 Harsh or abrasive detergents, and some solvents, can damage the display unit and information on the labels. CAUTION When using a Bird 5011, do not exceed 2 W average or 125 W peak power for 5 µs. Doing so will render the sensor inoperative. CAUTION +22 dBm max.
  • Page 5 UTILISATION TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L’INSTRUMENT. IMPIEGO QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA COME SPECIFICATO DAL PRODUTTORE LA PROZIONE DI SICUREZZA POTREBBE VENIRNE COMPROMESSA.
  • Page 6: About This Manual

    Bird Site Analyzer About This Manual This instruction book covers the Bird Site Analyzer Models SA-2000, SA-2000A, and SA-2500A. Changes to This Manual We have made every effort to ensure this manual is accurate. If you should discover any errors, or if you have suggestions for improving this manual, please send your comments to our Solon, Ohio factory.
  • Page 7 Utilities — Describes utilities to set up the Site Analyzer. Computer Software — Provides installation instructions and lists the features of the Bird Site Analyzer PC Software. Maintenance — Lists routine maintenance tasks for the Site Analyzer, as well as troubleshooting for common problems.
  • Page 8: Table Of Contents

    Table of Contents Safety Precautions ......... i About This Manual .
  • Page 9 Fault Location Mode........33 Setting the Frequency Span .
  • Page 10 Calibrating the Bird 5011 ....... . . 88 Utilities..........91 Adjust Date and Time .
  • Page 11: Introduction

    Data is graphed at 238 points across a user-specified frequency band or distance range. Transmitter systems are tested by using a Site Analyzer and a Bird power sensor to measure RF power. Data is displayed as power or match efficiency, depending on the sensor.
  • Page 12: Items Not Supplied

    Bird Site Analyzer Items Not Supplied Calibration Combination Bird 5010 Directional Power Sensor Bird 5011 Terminating Power Sensor Site Analyzer Features General Field replaceable Li-ion battery. Automatic power down conserves battery life. Antenna Test Rejects on-channel interfering signals to +13 dBm.
  • Page 13: Transmitter Test

    X and Y scales and units are user adjustable. Dual measurement markers and one difference marker. Measurement hold to temporarily store a trace. Printing capability. The Bird Site Analyzer is compatible with all printers that use HP PCL Level 3, including most HP inkjet printers. Fast swept measurement.
  • Page 14: Component Description

    Bird Site Analyzer Component Description...
  • Page 15 Introduction Activates the function described directly to the Soft Keys right of the key. Mode Keys Activates the mode menu. This menu allows the Mode user to select Measure Match, Fault Location, Measure Power, or Utilities modes. Activates the configuration menu for the current Config mode.
  • Page 16 Bird Site Analyzer Right Arrow During: Moves active marker right one point at a time. Marker Control As defined by the function description. All other times Up Arrow During: Increases the numeric value. Data Entry Moves the marker to the maximum trace value.
  • Page 17: Connection Description

    9600 baud, 8 data bits, 1 stop bit, no parity, and no handshake. Connect the Site Analyzer to a PC serial port or to Bird power sensors. Input for external power supplies. Plug DC Input either the ac power supply or the cigarette lighter adapter into the dc input.
  • Page 18: Display Description

    Bird Site Analyzer Display Description Measure Match Measure Match Mode Sample 000, ABC, 01-JAN-2001 12:01 Display Save/ Recall 1.70 Auto Scale Limit Test On 1.00 825.00 841.00 Calibration: Full M1: 835.20 MHz 1.64 Limit: 1.43 M2: 831.50 MHz 1.20 Print ∆M:...
  • Page 19 Introduction Graphic display of the measurement. Trace Indicates the calibration status. Calibration Indicator Indicates the limit value. Limit Line Value Displays FAIL if any part of the trace Limit Test Indicator exceeds the limit value. A horizontal line that graphically Limit Line displays the limit value.
  • Page 20 Bird Site Analyzer Measure Power Measure Power Mode Sample Sensor Connected Display Display 0.0 Watts PwrUnit Watts 600.00 MatchUnit Rtn Loss Refl 0.00 Watts Match 0.00 Loss(dB) Indicates the current status of power Sensor Status sensor. Identifies the measurement displayed on Primary Measurement the analog dial.
  • Page 21: Getting Started

    “LO”. When using an external power source, a power cord symbol replaces the battery gauge. The Bird Site Analyzer can be operated using an ac Adapters adapter or a 12 V automobile cigarette lighter adapter.
  • Page 22 Bird Site Analyzer Electronic Corporation SA-BATPAK CHARGER Battery 2.7A SA-BATPAK To charge the external battery pack: 1. Plug the automobile cigarette lighter adapter into the external battery pack on the side that says “To Battery”. Refer to the figure above.
  • Page 23: Power Up

    Model SA-2500A, 780 - 2500 MHz SN: 00000000 8 Feb 2002 Copyright © 1998 - 2002 by Bird Electronic Corporation Have the following system information ready before System Information you begin using the Site Analyzer: Frequency Span —...
  • Page 24 Bird Site Analyzer...
  • Page 25: Calibration

    Chapter 3 Calibration Calibration Accessories Calibration Combination including: One 50 ohm load One Open standard One Short standard Test Cable (optional) – Phase-stable for reliable, consistent results. NOTE: If a test cable is used to connect the Site Analyzer system under test, attach the standards to the end of the test cable during calibration.
  • Page 26 Bird Site Analyzer Press the CALIBRATE key. Attach the open standard. Press the OPEN key. Wait for a “beep” and for the trace to scroll before continuing. Attach the short standard. Press the SHORT key. Wait for a “beep” and for the trace to scroll before continuing.
  • Page 27: Measure Match Mode

    Chapter 4 Measure Match Mode This measurement verifies and monitors the match conditions in the antenna system at various frequencies. The results are shown on an x-y graph. Frequency is shown on the x-axis while return loss, cable loss, or VSWR is shown on the y-axis. Set the frequency and calibrate the Site Analyzer before taking any readings.
  • Page 28: Setting The Frequency

    Bird Site Analyzer Setting the Frequency Frequencies can be set manually or chosen from a list of presets. If the start, stop, center, or span is manually changed, the band will become “Custom”. NOTE: Changing the frequency settings will automatically turn calibration off. Always set the frequency before calibrating the unit.
  • Page 29: Band List

    Measure Match Mode Measure Match Mode Freq 99.99 Scroll to Start, Stop, Center, or Scale Span Limit Line 1.00 Cable Enter a new Loss 1500.00 2500.00 value Start: 1500.00 Center: 2000.00 MHz Stop: 2500.00 MHz Span: 1000.00 MHz Band List: Custom The band list pop-up menu contains frequency band Band List presets.
  • Page 30 Bird Site Analyzer Measure Match Mode Freq Display the band list and select a Band Start Stop Scale frequency <Full Band> 780.00 2500.00 band IS-553(Rx) 824300 849.00 IS-553(Tx) 869.00 894.00 Limit IS-54(Rx) 824.00 849.00 Line IS-54(Tx) 869.00 894.00 IS-95(Rx) 1850.00 1910.00...
  • Page 31: Setting The Scale & Unit Of Measure

    Measure Match Mode Setting the Scale & Unit of Measure The display scale can be set manually or by using Auto Scale. The display can have units of return loss [dB], cable loss [dB], or VSWR [ratio]. Measure Match Mode Press from Measure Freq...
  • Page 32: Auto Scale

    Bird Site Analyzer Auto Scale automatically sets the display scale so that Auto Scale the entire trace is displayed. Measure Match Mode Freq 1.70 Scroll to Auto Scale Scale Limit Line Activate Auto Scale 1.00 Cable Loss 780.00 2500.00 Min: 1.00...
  • Page 33: Setting Units

    Measure Match Mode Setting Units Measure Match Mode Freq 99.99 Scroll to Units Scale Limit Line 1.00 Cable Loss 780.00 2500.00 Min: 1.00 Auto Scale Max: 99.99 Units: VSWR Measure Match Mode Freq 99.99 Display the units list and select Scale VSWR, Cable Loss,...
  • Page 34 Bird Site Analyzer Measure Match Mode Freq Activate the selected unit Scale Limit Line -60.0 Cable Loss 780.00 2500.00 Min: -60.0 dB Auto Scale Max: 0.0 dB Units: Rtn Loss...
  • Page 35: Cable Loss Measurements

    Measure Match Mode Cable Loss Measurements To measure cable loss, the cable being tested should have an open on the far end. Measure Match Mode Cable Loss Mode Press from Freq Measure Match Mode Scale Limit Line -60.0 Press to Cable activate Loss...
  • Page 36: Limit Line

    Bird Site Analyzer Limit Line The limit line is an option that helps you see a failure. It appears as a horizontal line at the limit line value. Press from Measure Match Mode Measure Match Mode Freq Press to 99.99...
  • Page 37: Limit Test

    Measure Match Mode Limit Test Limit Test compares the trace to the limit line. Measure Match Mode Save/ Recall 1.70 Auto Press from Scale Measure Match Mode Limit Toggle Test On between Off, Audio, & On 1.00 HOLD 825.00 841.00 Calibration: Full Limit: 1.43 Print...
  • Page 38: Marker Adjustment

    Bird Site Analyzer Marker Adjustment Markers indicate VSWR to 0.01, or Return or Cable Loss to 0.1 dB. The exact values of M1, M2, and the delta marker (difference of M1 and M2) are displayed at the bottom of the screen.
  • Page 39 Measure Match Mode Move across the trace Measure Match Mode Active 1.70 Locate the highest point of the trace ∆M Single Locate the lowest point 1.00 of the trace 825.00 841.00 Calibration: Full M1: 835.20 MHz 1.64 M2: 831.50 MHz 1.20 ∆M: 3.70 MHz...
  • Page 40: Measurement Hold

    Bird Site Analyzer Measurement Hold Measurement Hold stops tracing at the end of the current measurement and displays the last measure- ment result. It is active when a lock appears in the indicator ball. Measure Match Mode Save/ Recall 1.70...
  • Page 41: Printing

    Measure Match Mode Printing The Bird Site Analyzer can print the information displayed on the screen to any HP Deskjet printer that supports the PCL Level 3 protocol. Measure Match Mode Save/ Recall 1.70 Auto Press from Scale Measure Match Mode...
  • Page 42 Bird Site Analyzer...
  • Page 43: Fault Location Mode

    Chapter 5 Fault Location Mode This measurement identifies the position of impedance discontinuities (faults) within the antenna/feeder system. The measurement results are displayed on an x-y graph. Distance from the Site Analyzer is shown on the x-axis, while relative magnitude of the discontinuity is shown on the y-axis.
  • Page 44: Setting The Frequency Span

    Bird Site Analyzer Setting the Frequency Span The frequency span and relative propagation velocity determine the maximum distance at which distance- to-fault measurements can be taken. For best results, use the table below to select the frequency span which matches the length and either the propagation velocity or dielectric constant of the cable under test.
  • Page 45 Fault Location Mode For example, for a cable with a dielectric constant of 2.778, the velocity percentage is 60%. If the cable is 100 feet long, then the Site Analyzer should be set to a frequency span of 700 MHz. The resolution of the graph will be 5 inches.
  • Page 46 Bird Site Analyzer Measure Match Mode Freq 99.99 Scroll to the Span value Scale Limit Line Enter the value from 1.00 Cable Table 1 Loss 1290.00 1990.00 Start: 1290.00 MHz Center: 1640.00 MHz Stop: 1990.00 MHz Span: 700.00 Band List: Custom...
  • Page 47: Setting The Cable Type

    Fault Location Mode Setting the Cable Type The cable type can be set manually or chosen from a list of presets. If the velocity of propagation or loss is manually changed, the cable will become “Custom”. NOTE: Changing the cable type or velocity of propagation will reset the distance scale to the maximum possible distance.
  • Page 48: Cable List

    Bird Site Analyzer Fault Location Mode Distance 99.99 Enter a new value Scale Limit Line 1.00 Smooth Meters 18.2 Vel. Prop.: 0.8800 Cable List: Custom Loss: 0.262 dB/m Cbl Type The cable list pop-up menu contains cable presets. Cable List Using a preset is quick, easy, and sets test parameters while eliminating a possible source of operator error.
  • Page 49 Fault Location Mode Fault Location Mode Distance 99.99 Activate the selected cable Scale Limit Line 1.00 Smooth 17.0 Meters Vel. Prop.: 0.8200 Cable List: 5062 Loss: 0.1447 dB/m Cbl Type...
  • Page 50: Setting The Distance

    Bird Site Analyzer Setting the Distance The maximum possible Stop Distance is determined by the frequency span. (See Table 1 on page 34) Press from Fault Location Mode Fault Location Mode Distance 99.99 Scale Press to display the distance Limit...
  • Page 51: Setting Units

    Fault Location Mode Setting Units Fault Location Mode Distance 99.99 Scroll to Units Scale Limit Line Display the units list and select 1.00 Meters or Smooth Feet 18.2 Meters Units Start: 0.0 m Units: Meters Meters Stop: 18.2 m Feet Cbl Type Fault Location Mode Distance...
  • Page 52: Setting The Scale & Unit Of Measure

    Bird Site Analyzer Setting the Scale & Unit of Measure The display scale can be set manually or by using Auto Scale. The display can have units of return loss [dB] or VSWR [ratio]. Press from Fault Location Mode Fault...
  • Page 53: Auto Scale

    Fault Location Mode Auto Scale automatically sets the display scale so that Auto Scale the entire trace is displayed. Fault Location Mode Distance 1.07 Scroll to Auto Scale Scale Limit Line Activate Auto Scale 1.00 Smooth 18.2 Meters Min: 1.00 Auto Scale Max: 99.99 Units: VSWR...
  • Page 54: Setting Units

    Bird Site Analyzer Setting Units Fault Location Mode Distance 99.99 Scroll to Units Scale Limit Line 1.00 Smooth Meters 18.2 Min: 1.00 Auto Scale Max: 99.99 Units: VSWR Cbl Type Fault Location Mode Distance 99.99 Display the units list and select...
  • Page 55 Fault Location Mode Fault Location Mode Distance Activate the selected unit Scale Limit Line -60.0 Smooth Meters 18.2 Min: -60.0 dB Auto Scale Max: 0.0 dB Units: Rtn Loss Cbl Type...
  • Page 56: Limit Line

    Bird Site Analyzer Limit Line The limit line is an option that helps you see a failure. It appears as a horizontal line at the limit line value. Press from Fault Location Mode Fault Location Mode Distance Press to 99.99...
  • Page 57: Limit Test

    Fault Location Mode Limit Test Limit Test compares the trace to the limit line. Fault Location Mode Save/ Recall 1.70 Auto Press from Scale Fault Location Mode Limit Test On Toggle between Off, Audio, & On 1.00 HOLD Meters 18.2 Calibration: Full Limit: 1.43 Print...
  • Page 58: Marker Adjustment

    Bird Site Analyzer Marker Adjustment Markers indicate VSWR to 0.01, or Return Loss to 0.1 dB. The exact values of M1, M2, and the delta marker (difference of M1 and M2) are displayed at the bottom of the screen. Press from...
  • Page 59 Fault Location Mode Move across the trace Fault Location Mode Active 1.70 Locate the highest point of the trace ∆M Single Locate the lowest point 1.00 of the trace Meters 18.2 Calibration: Full 13.1 m 1.32 8.4 m 1.20 ∆M: 4.7 m 0.12...
  • Page 60: Smooth

    Bird Site Analyzer Smooth Smooth is an option that will digitally average the displayed trace. Three levels of smooth are available. Fault Location Mode Distance 1.70 Press from Fault Location Scale Mode Limit Line 1.00 Smooth None Meters 10.0 Calibration: Full...
  • Page 61: Measurement Hold

    Fault Location Mode Measurement Hold Measurement Hold stops tracing at the end of the current measurement and displays the last measure- ment result. It is active when a lock appears in the indicator ball. Fault Location Mode Save/ Recall 1.70 Auto Press from Scale...
  • Page 62: Printing

    Bird Site Analyzer Printing The Bird Site Analyzer can print the information displayed on the screen to any HP Deskjet printer that supports the PCL Level 3 protocol. Fault Location Mode Save/ Recall 1.70 Auto Press from Scale Fault Location...
  • Page 63: Save And Recall

    Chapter 6 Save and Recall This chapter applies to Bird Site Analyzers with firmware from 1-Nov-2001 or later. If you have an older firmware version, see “Old Save and Recall” on page 71. Traces and setups can be stored in nonvolatile memory, along with descriptive labels and a time-date stamp.
  • Page 64: Save Trace

    Bird Site Analyzer Save Trace Save and Recall Save Description Date 001 015 ALPHA RX02 05-OCT-2001 11:26 Recall 05-OCT-2001 11:26 Press from Save and Recall Label The trace is saved Delete 2 of 2: Setup...
  • Page 65: Trace Label

    Save and Recall Trace Label Save and Recall Delete QuickText A D G J M P S V Y . Press from Save and Recall B E H K N Q T W Z , The Label menu is C F I L O R U X / displayed [Space] Shortcuts off.
  • Page 66 Bird Site Analyzer If the numeric shortcuts are off, press a number key Entering Numbers once to enter a number. If the numeric shortcuts are on, press the key twice. Save and Recall Delete Enter numbers using the QuickText A D G J M P S V Y .
  • Page 67 Save and Recall Changing Labels The label Save and Recall should be RX01, not TX01 Delete 016 BETA TX01 QuickText A D G J M P S V Y . B E H K N Q T W Z , C F I L O R U X / [Space] Shortcuts off.
  • Page 68 Bird Site Analyzer Deleting Labels Scroll to the Save and Recall text to be deleted Delete 016 TX01 QuickText A D G J M P S V Y . B E H K N Q T W Z , Press to...
  • Page 69 Save and Recall Saving Labels A complete Save and Recall label Delete 016 BETA TX01 QuickText A D G J M P S V Y . B E H K N Q T W Z , C F I L O R U X / [Space] Shortcuts off.
  • Page 70: Trace Label Quicktext

    Bird Site Analyzer Trace Label Quicktext Store commonly used label elements in the Quicktext to speed label entry. Save and Recall Select From Trace Label, scroll past the QuickText A D G J M P S V Y . Add to...
  • Page 71 Save and Recall Create Quicktext Enter text Save and Recall into the label, then activate Select Quicktext 016 BETA QuickText A D G J M P S V Y . Add to Qtext B E H K N Q T W Z , Delete C F I L O R U X / View...
  • Page 72 Bird Site Analyzer Quicktext Save and Recall Select Scroll to a Quicktext entry QuickText A D G J M P S V Y . Add to Qtext BETA B E H K N Q T W Z , ALPHA Delete...
  • Page 73 Save and Recall Delete Quicktext Save and Recall Select Scroll to a Quicktext entry QuickText A D G J M P S V Y . Add to Qtext BETA B E H K N Q T W Z , ALPHA Delete C F I L O R U X / View...
  • Page 74: Trace Label Config

    Bird Site Analyzer Trace Label Config Label Help provides instructions for labelling traces. Label Help Save and Recall Help Press from 016 BETA X01 Trace Label QuickText Use LEFT or RIGHT arrows to select column or A D G J M P S V Y .
  • Page 75: Save Setup

    Save and Recall Save Setup Saving the setup will save the following: Calibration Coeff. Start Frequency Center Frequency Limit Line Stop Frequency Span Frequency Scale Min Start Distance Units Scale Max Stop Distance Save and Recall Save Frequency Setup Temp Date 824-849 05-OCT-2001 09:01a...
  • Page 76: Recall Trace

    Bird Site Analyzer Recall Trace When a trace is recalled it is displayed along with the trace currently being measured. The frequency range will be changed to the recalled trace’s settings. To remove a recalled trace from the display, press the ESCAPE key from the Save and Recall screen.
  • Page 77: Recall Setup

    Save and Recall Recall Setup Save and Recall Press from Save and Save Recall Frequency Setup Temp Date The Setup 824-849 05-OCT-2001 09:01a menu is Recall 780-2500 23-OCT-2001 11:26a displayed Scroll to the Delete setup to be recalled Traces Measure Match Mode Save/ Recall 99.99...
  • Page 78: Delete Trace

    Bird Site Analyzer Delete Trace Save and Recall Save From Save Description Date and Recall, scroll to the 001 015 ALPHA RX02 05-OCT-2001 11:26 Recall trace to be 000 016 BETA TX01 05-OCT-2001 11:26 deleted Label Delete 2 of 2:...
  • Page 79: Delete Setup

    Save and Recall Delete Setup Save and Recall Press from Save and Save Recall Frequency Setup Temp Date The Setup 824-849 05-OCT-2001 09:01a menu is Recall 780-2500 23-OCT-2001 11:26a displayed Scroll to the Delete setup to be deleted Traces Save and Recall Save Frequency Setup...
  • Page 80 Bird Site Analyzer...
  • Page 81: Old Save And Recall

    Chapter 7 Old Save and Recall This chapter applies to Bird Site Analyzers with firmware from before 1-Nov-2001. If you have a more recent firmware version, see “Save and Recall” on page 53. Measurement data is stored in non-volatile memory.
  • Page 82: Label And Sort Trace

    Bird Site Analyzer Label and Sort Trace Press from Save and Recall The Label menu is displayed Press to display the Site setting Enter site number...
  • Page 83 Old Save and Recall Toggle between [Blank], Alpha, Beta, and Gamma Toggle between [Blank], Transmit, and Receive. Also displays the antenna number Enter antenna number Traces are automatically sorted in the order of Site, Sector, Antenna, and Date (with unlabeled traces at the top of the list) when the Done key is pressed.
  • Page 84 Bird Site Analyzer Press to sort traces Press to complete trace labeling...
  • Page 85: Save Setup

    Old Save and Recall Save Setup Five setups can be stored in the unit. Saving the setup will save the following: Calibration Coefficients Dielectric Start Frequency Stop Frequency Center Frequency Span Frequency Mode Scale Units Minimum Maximum Smoothing Limit Test Settings Cable Type Press from Save and...
  • Page 86: Recall Trace

    Bird Site Analyzer Recall Trace When a trace is recalled it is displayed along with the trace currently being measured. The frequency range will be changed to the recalled trace’s settings. To remove a recalled trace from the display, press the ESCAPE key from the Save and Recall screen.
  • Page 87: Recall Setup

    Old Save and Recall Recall Setup Press from Save and Recall The Setup menu is displayed Scroll to the setup to be recalled The setup is recalled...
  • Page 88: Delete Trace

    Bird Site Analyzer Delete Trace From Save and Recall, scroll to the trace to be deleted Press to delete the trace. Press again when asked “Are You Sure?”...
  • Page 89: Delete Setup

    Old Save and Recall Delete Setup Press from Save and Recall The Setup menu is displayed Scroll to the setup to be deleted Press to delete the setup. Press again when asked “Are You Sure?”...
  • Page 90 Bird Site Analyzer...
  • Page 91: Measure Power Mode

    An external sensor is required to measure RF power. Press from Select Mode CAUTION When using a Bird 5011, do not exceed 2 W average or 125 W peak power for 5 µs. Doing so will render the sensor inoperative.
  • Page 92: Connecting A Sensor

    Remote Power Sensor Port must be used. Connect the Bird Site Analyzer to power sensors using the built-in serial port, labeled “Remote Power Sensor” (see “Connection Description” on page 7). A 9 pin serial cable should be used.
  • Page 93: Setting The Full Scale Power

    Measure Power Mode Setting the Full Scale Power The Bird Site Analyzer will automatically set the full scale power for sensors that do not use elements. For element-based sensors, enter the power listed on the forward element. The reflected element’s power rating should be 10% of the forward element’s.
  • Page 94: Setting The Offset

    Bird Site Analyzer Setting the Offset To read unattenuated power when using a coupler or attenuator, enter (in dB) the attenuation or coupling factor. To convert percentages to dB, use the equation: Attenuation(dB) = 10 x Log [Attenuation(%) / 100]...
  • Page 95: Recall Setups

    Measure Power Mode Recall Setups Power measurement setups can be created using the Bird Site Analyzer PCTool Software, then stored in the Site Analyzer’s nonvolatile memory. These setups store both the offset and full scale power. Measure Power Mode Press from...
  • Page 96: Choosing The Displayed Measurement

    Bird Site Analyzer Choosing the Displayed Measurement The primary display shows one value on the large numerical display and the dial. Other measurements are displayed numerically below the dial. NOTE: When using a terminating power sensor, only forward power can be measured. Reflected power and match will not be displayed.
  • Page 97: Setting Units

    Measure Power Mode Setting Units Toggle between Watts and Measure Power Mode Sensor Connected Display -30.00 dBm PwrUnit 58.00 Toggle MatchUnit between VSWR VSWR, Return Loss, and Match Efficiency Refl -30.00 dBm Match 2.00 VSWR...
  • Page 98: Calibrating The Bird 5011

    Bird Site Analyzer Calibrating the Bird 5011 CAUTION When using a Bird 5011, do not exceed 2 W average or 125 W peak power for 5 µs. Doing so will render the sensor inoperative. The Bird 5011 is a highly accurate terminating power sensor.
  • Page 99 Measure Power Mode Measure Power Mode Configuration Screen Recall Setups Press to the 0.00 mWatts warning Calibrate 5011 12.00 WARNING Do not disconnect sensor until cal is complete. Offset 0.00 FullScale 0.01 Watts Wait for Measure Power Mode calibration to finish Configuration Screen Recall Setups...
  • Page 100 Bird Site Analyzer...
  • Page 101: Utilities

    Chapter 9 Utilities These utilities adjust the Bird Site Analyzer’s date and time, return the unit to default settings, or check the printer status. Utility Mode System Press the Mode key Model SA-2500A, 780 - 2500 MHz Printer SN: 00000000...
  • Page 102: Adjust Date And Time

    Bird Site Analyzer Adjust Date and Time Utility Mode System Model SA-2500A, 780 - 2500 MHz Printer SN: 00000000 Press to display the 8 Feb 2002 system settings FM Mod Current Date / Time: 12 - - 2001 12:05 p...
  • Page 103: Return To Defaults

    Utilities Return to Defaults All settings will be returned to the factory presets. Saved traces and setups will not be affected. This function should be used after a unit failure and on first power up. Utility Mode System Model SA-2500A, 780 - 2500 MHz Printer SN: 00000000 Press to...
  • Page 104: Fm Modulation

    Bird Site Analyzer FM Modulation The Bird Site Analyzer uses FM modulation to improve its immunity to interfering signals. Under certain circumstances, such as while making cable loss or other high reflection measurements with long cables, FM modulation may reduce signal quality. In that case, the modulation should be turned off.
  • Page 105: Printer

    Utilities Printer The Bird Site Analyzer is compatible all printers that use HP PCL Level 3, including most HP printers. When a printer error occurs, this screen will display an error message describing the printer problem. Utility Mode System Model SA-2500A, 780 - 2500 MHz...
  • Page 106 Bird Site Analyzer...
  • Page 107: Computer Software

    Chapter 10 Computer Software The Bird Site Analyzer Software is designed to help you use the Site Analyzer more effectively. The software enables use of a PC for archiving and analyzing measurement data. Features Multiple Document Interface - allows any number of trace documents to be open simultaneously.
  • Page 108 Bird Site Analyzer...
  • Page 109: Maintenance

    Chapter 11 Maintenance Cleaning CAUTION Harsh or abrasive detergents, and some solvents, can damage the display unit and labels. Only clean the Site Analyzer with a soft cloth dampened with mild detergent and water. Do not use any other type of cleaning solution. Charging the Battery The internal battery pack is charged while the Site Analyzer is connected to the ac adapter or the...
  • Page 110: Troubleshooting

    Bird Site Analyzer Troubleshooting Any service procedure not covered in this manual should be referred to an authorized service facility. Locate the problem, review the possible causes, and perform the action listed. If the problem is not corrected, give us a call or return the unit for service.
  • Page 111 Maintenance Problem Possible Cause Possible Correction Keys do not respond Unit is “Locked Up” Turn the unit off and then back on. Perform a full system initialization. Disconnect the AC adapter, hold down the Contrast key and turn the unit off. Turn the unit back on.
  • Page 112: Battery Replacement

    Bird Site Analyzer Battery Replacement Lay the Site Analyzer, display side down, on a clean surface. Remove the four screws and the battery cover. Refer to the figure below. NOTE: Check the direction of the notch in the Battery Cover. It is important to replace the cover the same way it was removed.
  • Page 113: Unit Reset

    Maintenance Unit Reset Lay the Site Analyzer, display side down, on a clean surface. Remove the battery. Follow the instructions in “Battery Replacement”, page 102. Remove the label covering the Reset and EPROM access. Refer to the figure on the previous page. Press the Reset button with a nonconductive instrument.
  • Page 114: Flash Rom Upgrade (Sa-2500A Only)

    Bird Site Analyzer Flash ROM Upgrade (SA-2500A Only) The SA-2500A uses flash-programmable ROM. It is not necessary to replace the EPROM to upgrade the firmware. NOTE: The update process will erase the memory of the Site Analyzer, including all saved traces and setups.
  • Page 115: Eprom Replacement

    Maintenance EPROM Replacement The EPROM should not need to be replaced by the user. If the EPROM does ever need to be replaced, it should only be replaced by qualified service personnel. NOTE: The Model SA-2500A uses flash- programmable ROM. It is not necessary to replace the EPROM to upgrade the firmware.
  • Page 116: Customer Service

    Bird Site Analyzer Customer Service If you need to return the unit for any reason, contact the Bird Service Center for a return authorization. All instruments returned must be shipped prepaid and to the attention of Bird Service Center. Bird Service Center...
  • Page 117: Specifications

    Maintenance Specifications Frequency Characteristics Frequency Range: Model SA-2000 806 MHz to 2000 MHz Model SA-2000A 806 MHz to 2300 MHz Model SA-2500A 780 MHz to 2500 MHz Frequency Resolution: 50 kHz Frequency Accuracy: ± 150 kHz (75 ppm) Number of Points: Measurement Range and Resolution Range Resolution...
  • Page 118 Same as for Return Loss Measurements 0.20 to 1.00, or 1.00 to 99 percent Range (new specs) Power Measure Mode Function: Displays power from Bird power sensors, VSWR alarm and BPM (specifications determined by sensor) General 300 traces in fundamental data format Data Storage stored in non-volatile memory.
  • Page 119 Maintenance Printer Interface Port Connector: Female DB-25, compatible with PC parallel port. Compatibility: HP Deskjet printers with PCL Level 3 protocol Power Requirements Internal: Lithium-ion rechargeable battery 3 hours minimum operating time External DC: 9 to 16 Vdc External AC: 90 to 264 Vac @ 45 to 66 Hz Physical Specifications Dimensions:...
  • Page 120 Bird Site Analyzer Measurement Uncertainty (Return Loss after one refresh)
  • Page 121: Optional Equipment Available

    Maintenance Optional Equipment Available Precision Open/Short/Load Combination: Connector Type Part Number Male N CAL-MN-B Female N CAL-FN-B Male 7/16 DIN CAL-ME-B Female 7/16 DIN CAL-FE-B Armored (phase stable) Test Cables: Connector Types Length Part Number Male N to Male N 1.5 meter TC-MNMN-1.5 Male N to Male N...
  • Page 122 Power Sensor for Thruline power measurement. Requires two Sensor DPM elements. Refer to DPM Element Guide, P/N 871-DPM-019-901, Elements for a complete list of Bird DPM elements. P/N: 5011 Terminating Power Sensor for terminating power measurement. Sensor Terminating Power Sensor Accessories:...
  • Page 123 Limited Warranty All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of two (2) years, unless otherwise specified, from date of shipment and to conform to applicable specifications, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be other than as warranted;...
  • Page 124 Site Analyzer Models: SA-2500A The undersigned hereby declares, on behalf of Bird Electronic Corporation of Cleveland, Ohio, that the above-referenced products, to which this declaration relates, are in conformance with the provisions of the following standards. European Standard EN 55011,1998 - Conducted & Radiated Emissions...

This manual is also suitable for:

Site analyzer sa-2500aSite analyzer sa-2000a

Table of Contents