4.6 Data Output RS232 inter face is recommended. And can be connected
to printer or computer
5. Displaying Information
5.1 Display window F----1 data beyond scope
F----2 AC/DC conversions
F----3 buttons
F----5 sensor faults
F----L lower zero position of sensor signal
F----H overload
C----F calibration error, instability or not at zero point during calibrating
C----H calibration error, items on the pan or at higher zero point
during calibrating
E------ Storage calibration data error, re-calibration is necessary
5.2 Indicator
kg(g) lightening with stable data, the weighting units is kg or
lb
lightening with stable data, the weighting units is lbr
pcs
lightening with stable data, and counting situation
Lower voltage lightening in lower voltage with warning. Charge is
needed, otherwise can only last 1—2 hours. Balance is still working during
charging.
Charge lightening during charging. One single charging needs 8 hours
Full charge slowly lightening at beginning of charging, and turning
off after full charging
6. Settings
Start the unit by holding down the "CAL" key till it comes to "F----3"
and you are in the setting mode. And pressing "CAL" can change the par-
ameters; pressing "TARE" can change the values of parameters.
6.1 C1------ sensitivity setting 0 1 2 3 4, the larger number is, the less
sensible but the better stable. Default setting is 2 or 1
6.2 C2------ filtering strength setting 0 1 2 3, the larger number is, the
lower reaction speed, but the better stable. Default setting is 2 or 1
5
6.3 C3------ Baud rate setting 2(600) 3(1200) 4(2400) 5(4800) 6(9600)
7(1200). Default setting is 2
6.4 C4------ communication setting. "TARE" changes the lower unit.
"COUNT" changes the higher units. This data is the first data during com-
munication. Default setting is 27.
6.5 C5------ units change. 0 is for units changing and 1 for printing.
Default setting is 0. Re-press "CAL" after setting. Then the balance is auto
calibrated to 0.
Start the unit by holding down the "TARE" key till it comes to "F----3".
All the parameters can be reset to default settings. Re-calibration before
using
7. Serial communication
Working way: 8 units data, 1 unit data, ending unit without calibration
Receiving data:
Signal+orders Print 70H (ACSII code p)
Calibration 71H (ACSII code q)
Accumulation 72H (ACSII code r)
Units change 73H (ACSII code s)
Tare 74H (ACSII code t)
Sending data
12345.6□kg↓←14 units
e.g. VB communication process for C3---2 C4—27 (Default settings)
MSComm1.Settings ="600,n,8,1"
MSComm1.Output= Chr(&H1B) + Chr(&H70)' sending print order
'MSComm1.Output= Chr(27) + Chr(112)
'MSComm1.Output= Chr(27)+ "p"
Do
DoEvents
Loop Unti MSComm1. InBufferCount = 14
a = MSComm1.Input
Print a