Error Messages - Fireco Deafgard Operating Manual

Fire and wake-up alarm
Hide thumbs Also See for Deafgard:
Table of Contents

Advertisement

Error Messages

ER01 – PAD
This may be shown on the LCD as 'ER01' or 'PAD'. The Deafgard will
regularly check the pillow pad to ensure it is present.
Resolution: Remove Pillow Pad from Cradle and then remove batteries wait
20 seconds and re-insert batteries, perform self test by pressing
"Mute/Snooze" button. If this does not resolve the error please contact
Technical Advice line.
ER02 – Motor Jam
Whenever the Deafgard performs a self test routine (once every 24Hrs, or on
demand) the operation of the pillow pad motor is checked. The motor is
turned on and then the battery voltage is checked. If the voltage drop is too
great it is determined that the motor is drawing too much current and has
stalled. A Motor Jam error code will then be displayed.
Resolution: Depress and hold Button 4 MUTE/SNOOZE until 'TEST' is
shown on Display (after approximately 4 seconds).
a. 'TEST' should be shown on Display.
b. Both Strobes should be flashing.
c. The 'bleep' should be continuous.
d. The LED should flash GREEN once every 15 seconds.
e. The Pillow pad should be vibrating.
After 10 seconds the Test should stop and:
a. Display should show 0:01 or similar, dependent upon time elapsed.
b. Both Strobes should be Off.
c. No 'bleep' should be heard.
d. The LED should flash GREEN once every 15 seconds.