Error which has
Possible causes
occurred
Cuff does not inflate
or does not inflate
fast enough. This can,
among other things,
be due to a too loosely
applied cuff or to
movement.
Cuff loses air/has a leak.
Air release during the
measuring procedure is
too fast or too slow. The
cuff could have become
undone or loosened
itself. A movement dur-
ing the measuring proce-
dure is also a possibility.
The pressure in cuff
exceeds 300 mmHg.
Therefore, an automatic
fall in pressure occurs.
If the battery symbol
flashes, the batteries are
almost dead. Only a few
more measurements are
possible.
Remedy
Apply the cuff so that it
fits snugly around the
wrist.
Return the device to
the designated service
address.
Check the correct
position of the cuff. Do
not move during the
measuring procedure.
Please rest for at least
a minute and take the
measurement again.
Keep new batteries of
the same type handy
(type AA LR03).
170
Error which has
Possible causes
occurred
If the battery symbol is
permanently illuminated
the batteries are dead
and have to be
replaced.
Implausible
Device not positioned at
measured values.
the level of the heart.
Cuff too big or too
small.
Cuff placed on top of
clothing.
Rolled-up clothing
impedes blood
circulation.
Cuff wrongly applied.
Cuff was not correctly
inflated.
English
Remedy
Insert new batteries of
the same type (type AA
LR03).
Position the wrist at the
level of the heart, and
repeat the measurement.
Use an upper arm blood
pressure monitor.
Apply cuff on the naked
skin.
Wear loose clothing.
Make sure that rolled-up
sleeves do not impair
circulation in the arm.
Take note of the
instructions and images
showing how to apply
the cuff correctly to the
wrist.
Check the correct
position of the cuff.