Omron RS8 Instruction Manual page 2

Wrist blood pressure monitor

Advertisement

HEM-6310F-E.book Page 2 Monday, March 2, 2015 4:39 PM
To View the Readings Stored in Memory
1.
Press the
button, while the average value is
displayed.
The Memory number
appears for a second
before the pulse rate is
displayed. The newest
Alternating
set is numbered "1".
date/time
display
2.
Press the
button repeatedly to view the
readings stored in memory.
To View the Weekly Averages
This unit calculates and displays weekly averages for the
measurements taken in the morning and evening within 8
weeks for each user.
Note: The week begins Sunday at 2:00.
1.
Select your User ID.
2.
Press the
button once.
The morning average for the current
week "THIS WEEK" appears on the
display.
3.
Press the
button again.
The evening average for the current
week "THIS WEEK" appears on the
display.
4.
Continue to press the
button to display the
previous weeks.
The unit displays "-1 WEEK" for the previous week to
"-7 WEEK" for the oldest set of averages.
Note: If there are not enough
measurements taken in a week
to calculate the average the
display will appear as
illustrated.
This week
This week
Last week
Morning average
Evening average
Morning average
Last week
Evening average
7 weeks ago
7 weeks ago
Evening average
Morning average
5.
Press the START/STOP button to turn off the
monitor.
Morning Weekly Average
4.1 The Icons and Error Messages
This is the average for the measurements taken during the
morning (4:00 - 11:59) between Sunday and Saturday. An
Error Display
average for each day is calculated for up to three
measurements taken within 10 minutes of the first
measurement of the morning.
Measurements of the morning
Sunday
Monday
Tuesday
Wednesday
Thursday
Friday
Saturday
4:00
First measurement of
Measurement
the morning
Measurement
Up to 3 measurements
Measurement
Within 10 minutes
Morning Blood Pressure
Measurement
Measurement
Measurement
11:59
Evening Weekly Average
This is the average for the measurements taken during the
evening (19:00 - 1:59) between Sunday and Saturday. An
average for each day is calculated for up to three
measurements taken within 10 minutes of the last
measurement of the evening.
Measurements of the evening
Sunday
Monday
Tuesday
Wednesday
Thursday
Friday
Saturday
19:00
Measurement
Measurement
Within 10 minutes
Measurement
Measurement
Measurement
Up to 3 measurements
Last measurement of
Measurement
the evening
Evening Blood Pressure
1:59
Note: The irregular heartbeat symbol (
The Blood Pressure Trend Graph
4.2 Troubleshooting
This unit also displays the Blood Pressure Trend Graph which
shows the weekly averages for the Systolic Blood Pressure
(SYS) taken in the morning and evening within 8 weeks for
each user. The graph is displayed 4 weeks averages so that
you can see the continuous blood pressure change.
Systolic blood
pressure
Note:It flashes to show
above169 mmHg
which week
average value is
152 - 168 mmHg
displayed.
The reading is
136 - 151 mmHg
extremely low
(or high).
119 - 135 mmHg
118 mmHg or less
3 weeks ago
2 weeks ago
Last week
This week
7 weeks ago
6 weeks ago
5 weeks ago
4 weeks ago
This week - 3 weeks ago
Wrist cuff pressure
does not rise.
This week
This week
Wrist cuff deflates
morning SYS average
evening SYS average
too soon.
118 mmHg
140 mmHg
4 weeks ago - 7 weeks ago
The blood pressure is different
each time. The reading is
extremely low (or high).
7 weeks ago
7 weeks ago
The unit loses
morning SYS average
evening SYS average
power during
112 mmHg
138 mmHg
measurement.
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by
Nothing happens
User ID.
when you press
When the memory symbol (
) appears, first press
the buttons.
the
button. Then while holding it down, press the
START/STOP button simultaneously for about 2 - 3
seconds.
Other problems.
First
Second
4.3 Transferring the Data to the Online
Note: You cannot partially delete the stored readings.
All values for the user you select will be deleted.
To begin using the software for the first time, please
visit www.omron-healthcare.com for the initial set-up
4. Troubleshooting and Maintenance
instructions.
Cause
Remedy
4.4 Maintenance
Remove the wrist cuff. Wait
To protect your unit from damage, please observe the
2 - 3 minutes and then take
following:
Irregular heartbeats
another measurement.
are detected.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Movement during
Carefully read and repeat the
measurement.
steps in section 3.3.
You should replace them with
The batteries are
new ones ahead of time.
Blink
low.
Refer to section 2.1.
You should replace them with
The batteries are
new ones at once.
Lit
exhausted.
Refer to section 2.1.
Wrist cuff not
Apply the wrist cuff correctly.
applied correctly.
Refer to section 3.1.
Repeat Measurement while
remaining still and refraining
Movement during
from talking during the
measurement.
measurement.
Refer to section 3.3.
Apply the wrist cuff correctly
Wrist cuff not
and repeat measurement
applied correctly, or
while remaining still and
movement during
refraining from taking during
measurement.
the measurement.
Refer to section 3.1 and 3.3.
Arm position
Remain still until the
changed during
measurement is complete.
measurement.
Refer to section 3.3.
Contact your OMRON retail
Device error.
outlet or distributor.
) and the movement error
symbol (
) may also be displayed with error messages.
Do not store the unit in the following situations:
Problem
Cause
Remedy
Measure while in the
The wrist cuff is not
correct posture.
at heart level.
Refer to section 3.2.
The cuff is not
Wrap the cuff correctly.
Calibration and Service
wrapped snugly
Refer to section 3.1.
around the wrist.
The arms and
Relax and try taking the
shoulders are
measurement again.
tense.
Refer to section 3.3.
Movement or
Remain still and do not
talking during
talk during Measurement.
measurement.
Refer to section 3.3.
5. Technical Data
Air is leaking from
Consult your OMRON
Product Description
the wrist cuff.
retail outlet or distributor.
Model
Display
Apply the cuff correctly so
Measurement Method
The wrist cuff is
that it is firmly wrapped
loose.
around the wrist. Refer to
Transmission Method
section 3.1.
Measurement Range
Blood pressure readings
constantly vary with time
Accuracy
of day and how relaxed
you are. Take several
Inflation
deep breaths and try to
remain relaxed before
Deflation
taking a measurement.
Memory
Power Source
Replace the batteries with
The batteries are
new ones.
Battery Life
exhausted.
Refer to section 2.1.
Applied Part
The batteries have
Insert the batteries with
been inserted
the correct (+/-) polarity.
Protection Against
incorrectly.
Refer to section 2.1.
Electric Shock
Operating temperature/
• Press the START/STOP button and repeat
Humidity
measurement.
Storage temperature/
• If the problem continues, try replacing the
Humidity/Air pressure
batteries with new ones.
If this still does not solve the problem, contact
Console Weight
your OMRON retail outlet or distributor.
Outer Dimensions
Measurable
circumference
Solution
Cuff Material
Package Content
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC
• Do not subject the main unit and the cuff to
(Medical Device Directive).
extreme temperatures, humidity, moisture or
• This blood pressure monitor is designed according to the European
direct sunlight.
Standard EN1060, Non-invasive sphygmomanometers Part 1:
General Requirements and Part 3: Supplementary requirements
• Do not disassemble the unit.
for electromechanical blood pressure measuring systems.
• Do not subject the unit to strong shocks or
• This OMRON product is produced under the strict quality system of
vibrations (for example, dropping the unit on the
OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
floor).
OMRON blood pressure monitors, which is the Pressure Sensor, is
• Do not use volatile liquids to clean the main unit.
produced in Japan.
• Do not wash the cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to
clean the cuff.
• Do not carry out repairs of any kind yourself. If a
defect occurs, consult your OMRON retail outlet
or distributor as mentioned on the packaging.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and neutral soap to
clean the cuff.
• Keep the unit in its storage case when not in
use.
• Fold the cuff into the storage case.
Important information regarding Electro Magnetic
Compatibility (EMC)
• If the unit is wet.
• Locations exposed to extreme temperatures,
With the increased number of electronic devices such as PC's and
humidity, direct sunlight, dust or corrosive
mobile (cellular) telephones, medical devices in use may be
susceptible to electromagnetic interference from other devices.
vapours.
Electromagnetic interference may result in incorrect operation of
• Locations exposed to vibrations, shocks or
the medical device and create a potentially unsafe situation.
where it will be at an angle.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic
Compatibility) with the aim to prevent unsafe product situations, the
• The accuracy of this blood pressure monitor has
EN60601-1-2:2007 standard has been implemented. This standard
been carefully tested and is designed for a long
defines the levels of immunity to electromagnetic interferences as
service life.
well as maximum levels of electromagnetic emissions for medical
devices.
• It is generally recommended to have the unit
inspected every two years to ensure correct
This medical device manufactured by OMRON HEALTHCARE
functioning and accuracy. Please consult your
conforms to this EN60601-1-2:2007 standard for both immunity and
emissions.
authorised OMRON dealer or the OMRON
Nevertheless, special precautions need to be observed:
Customer Service at the address given on the
• Do not use mobile (cellular) telephones and other devices, which
packaging or attached literature.
generate strong electrical or electromagnetic fields, near the
medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is
Wrist Blood Pressure Monitor
available at OMRON HEALTHCARE EUROPE at the address
OMRON RS8 (HEM-6310F-E)
mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
LCD Digital Display
Oscillometric method
Correct Disposal of This Product
NFC-F (conforming to ISO/IEC)
(Waste Electrical & Electronic Equipment)
Pressure: 0 to 299 mmHg
This marking shown on the product or its
Pulse: 40 to 180 beats/min.
literature, indicates that it should not be
Pressure: ±3 mmHg
disposed of, with other household wastes at the
Pulse: ±5% of display reading
end of its working life.
To prevent possible harm to the environment or human health from
Automatic inflation by pump
uncontrolled waste disposal, please separate this product from
Automatic rapid deflation
other types of wastes and recycle it responsibly to promote the
sustainable reuse of material resources.
100 Measurements with date and time for
each user (A and B)
Household users should contact either the retailer where they
2 x 1.5V (LR03, AAA alkaline batteries)
purchased this product, or their local government office, for details
of where and how they can return this item for environmentally safe
Approx. 300 measurements with new
recycling.
alkaline batteries at a room temperature of
Business users should contact their supplier and check the terms
23°C
and conditions of the purchase contract. This product should not be
mixed with other commercial wastes for disposal.
= Type B
Manufacturer
Internally powered ME equipment
+10 to +40°C / Maximum: 30 to 85% RH
EU-representative
-20 to +60°C / Maximum: 10 to 95% RH /
700 to 1060 hPa
Approximately 80 g without batteries
Production Facility
Approximately 89 (w) mm x 61 (h) mm x
13 (d) mm (without the wrist cuff)
Approximately 13.5 to 21.5 cm
Subsidiaries
Nylon and polyester
Main unit, storage case, battery set,
instruction manual, guarantee card, blood
pressure pass
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, KYOTO,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp,
THE NETHERLANDS
www.omron-healthcare.com
OMRON DALIAN Co., Ltd.
No. 3, Song Jiang Road,
Economic and Technical Development Zone,
Dalian 116600, China
OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG,
UK
www.omron-healthcare.co.uk
OMRON MEDIZINTECHNIK
HANDELSGESELLSCHAFT mbH
Gottlieb-Daimler-Strasse 10, 68165 Mannheim,
GERMANY
www.omron-healthcare.de
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
Cedex, FRANCE
www.omron-healthcare.fr
Made in China