End Fitting Screen
To configure the End Fitting screen:
1. Use the controls to adjust the Beep
Frequency and Beep Intensity, if required.
2. Check on or off for low battery warnings,
program beeps, ideal volume indicator
beep and OnBoard control VC beeps by
selecting and deselecting the appropriate
check boxes.
3. Select/Deselect the Data Logging
checkbox to enable/disable Data Logging.
4. Activate the Startup Delay option by
moving the slider to the appropriate
position. Choose from: No Delay, 10
seconds or 15 seconds. The default
position is No Delay.
To demonstrate the program beeps, low
battery warning, ideal volume indicator or
OnBoard control beeps, click the musical
notes in the left DEMONSTRATE BEEPS menu.
Note: If Easy-t is selected as a program, there are two different buttons to demonstrate the
signal. The first button demonstrates the signal when the client accesses the telephone
program manually. The second Easy-t button demonstrates the signal that is heard when
the telephone program activates automatically.
Save and Exit
To end the programming session and save the information to the database and the hearing
aid(s):
1. Click Exit U:fit.
2. Select either the Save or Don't Save options.
a. If the Save option is selected, choose whether to Save to Database and/or Save to
Hearing Aid.
3. Click OK.
To save without exiting U:fit, click Save... in the bottom right corner of the End Fitting
screen. Save... closes the client record without exiting U:fit.