AC voltages 100-240 VAC (47-63 Hz). The 50MS-192 is also AC current protected by use of a 2 Amp “Slo-Blo” AC fuse. The fuse is field replaceable in the event of any failure to the fuse. The fuse itself is a 5x20 mm “Slo-Blo” type fuse and can be ordered through JFW or directly from Littelfuse.
4 - Command Set The command set for both RS-232 and Ethernet is as follows: Set Attenuator Syntax: SAx y <CR> x = attenuator number to control y = attenuation value to set the attenuator to <CR> = carriage return Description: This command sets attenuator “x”...
<CR> = carriage return Description: This command returns the identification information for this system and is followed by a carriage return and a line feed. It will list JFW Industries Inc, followed by the JFW model number and the firmware revision level. Examples: ID <CR>...
To configure the Ethernet, connect the provided “Null Modem” cable from the “Ethernet Configuration Port” on the 50MS-192 to a COM port on an available PC. See section 7 of this manual for specific instructions on Ethernet configuration. The easiest method of verifying the unit is working correctly after your changes is to use the JFW provided test program.
Open up a terminal session through your computer’s COM port using a program like HyperTerminal. You must use the Ethernet Configuration Cable to make the physical connection from your COM port to the “Ethernet Config. Port” on the rear of the 50MS-192. The terminal session should use the following COM port settings:...
RS-232 Port and to configure the Ethernet connection. This cable is DE- 9P to DE-9S and is the “Null Modem” type. The female connector will plug into the serial port on most PC’s, and the male connector will connect to 50MS-192. This cable can also be bought from L-Com (L-Com #CSNULL9MF-10) 10 - RS-232 Port Settings When sending commands to the 50MS-192, your computer’s RS-232 port settings must be...
Run the executable file 50MS-192.EXE to start the program. The test program can be used to verify functionality of the 50MS-192. A picture of the test program is shown below. Before exiting the program, it is a good idea to close any connections you have made while using the software.
12 - Error Code List Error1 Command is not formatted incorrectly. This error occurs if characters other than ID, SI, RI, SA or RA appear in the buffer. Error2 Attenuator address or Input port value out of range. Valid Attenuator addresses are 1-3. Valid Input port numbers are 1-24.
13 - TROUBLESHOOTING Initialization Time After the 50MS-192 has been turned on, it will take at least 8 seconds for the unit to completely initialize itself. If you send remote commands while it is still initializing, the command will not be executed.