Electromagnetic Compatibility Information

Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emissions
The IH30 is intended for use in the electromagnetic environment specified below. The customer or the
user of the IH30 should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The IH30 is intended for use in the electromagnetic environment specified below. The customer or the
user of the IH30 should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transi-
ent/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test level Compliance level
± 6 kV contact
± 8 kV air
±2 kV for power sup-
ply lines
±1 kV line(s) and neu-
tral
<5 % U
(>95 % dip
T
in U
) for 0,5 cycle 40
T
% U
(60 % dip in U
)
T
T
for 5 cycles 70 % U
T
(30 % dip in U
) for 25
T
cycles <5 % U
(>95
T
% dip in U
) for 5s
T
3 A/m
Electromagnetic environment – guidance
The IH30 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The IH30 is suitable for use in all establishments, inclu-
ding domestic establishments and those directly con-
nected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
± 6 kV contact
± 8 kV air
±2 kV for power sup-
ply lines
±1 kV line(s) and neu-
tral
<5 % U
(>95 % dip
T
in U
) for 0,5 cycle 40
T
% U
(60 % dip in U
)
T
T
for 5 cycles 70 % U
T
(30 % dip in U
) for 25
T
cycles <5 % U
(>95 %
T
dip in U
) for 5s
T
3 A/m
66
Electromagnetic environment
– guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the IH30 requires continued
operation during power mains
interruptions, it is recommen-
ded that the IH30 be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or hos-
pital environment.